Troubleshooting
Troubleshooting information for the various circuits is summarized below. See the
component layout drawings at the end of Section 6 for test point locations.
Display board checks
If the front panel display tests indicate that there is a problem on the display board, use Table
4-1. See principles of operation for display circuit theory.
Table 4-1
Display board checks
Step Item/component
Required condition
Remarks
1
2
3
4
5
6
7
Front panel test
P1005, pin 5
P1005, pin 9
U902, pin 1
U902, pin 43
U902, pin 32
U902, pin 33
Verify that all segments operate.
+5V ±5%
+37V ±5%
Goes low briefly on power up,
then goes high.
4MHz square wave.
Pulse train every 1ms.
Brief pulse train when front
panel key is pressed.
Use front panel display test.
D5V supply.
D37V supply.
Microcontroller RESET.
Controller 4MHz clock.
Control from main processor.
Key down data sent to main
processor.
Troubleshooting
4-13
2410-902-01.book Page 13 Monday, November 7, 2005 2:49 PM
Содержание Keithley SourceMeter 2410
Страница 11: ...1 Performance Verification 2410 902 01 book Page 1 Monday November 7 2005 2 49 PM ...
Страница 29: ...2 Calibration 2410 902 01 book Page 1 Monday November 7 2005 2 49 PM ...
Страница 53: ...3 Routine Maintenance 2410 902 01 book Page 1 Monday November 7 2005 2 49 PM ...
Страница 56: ...4 Troubleshooting 2410 902 01 book Page 1 Monday November 7 2005 2 49 PM ...
Страница 74: ...5 Disassembly 2410 902 01 book Page 1 Monday November 7 2005 2 49 PM ...
Страница 84: ...6 Replaceable Parts 2410 902 01 book Page 1 Monday November 7 2005 2 49 PM ...
Страница 100: ...A Specifications 2410 902 01 book Page 1 Monday November 7 2005 2 49 PM ...
Страница 102: ...B Command Reference 2410 902 01 book Page 1 Monday November 7 2005 2 49 PM ...
Страница 116: ...C Calibration Programs 2410 902 01 book Page 1 Monday November 7 2005 2 49 PM ...