Section 6: On-state characterization of a power MOSFET
Model 8010 High Power Device Test Fixture User's Manual
6-8
8010-900-01 Rev. C / March 2017
Set up communication
The communication setup is illustrated in the following diagram. GPIB is used as an example, but this
application can be run using any of the supported communication interfaces for the instruments. A
TSP-Link
®
connection enables communication between two instruments. Commands for the
Model 2651 SMU #2 (node #2) and the Series 2600 SMU #3 (node #3) are sent over the TSP-Link
interface.
Figure 63: GPIB communication example for Rds(on) sweep
To set the TSP-Link node number using the front-panel interface:
1. Press the
MENU
key.
2. Select
TSPLink
.
3. Select
NODE
.
4. Use the navigation wheel to adjust the node number.
5. Press the
ENTER
key to save the TSP-Link node number.
On the Model 2651 SMU #1 (TSP-Link node #1), perform a TSP-Link reset to update it with the linked
instruments:
1. Press the
MENU
key.
2. Select
TSPLink
.
3. Select
RESET
.