Model 8010 High Power Device Test Fixture User's Manual
Section 3: Working with devices and instruments
8010-900-01 Rev. C / March 2017
3-33
If there is no cable connected to the HI and SHI connections of the 2657, use the discharge probe to
contact the center conductor of the HI and SHI BNC connectors. Hold there for at least one second
to safely discharge system capacitance.
Figure 46: 8020-DP contact HI BNC connector
6. Use the probe to contact the 2657 GUARD banana jack. Hold for at least one second to safely
discharge system capacitance.
Figure 47: 8020-DP contact GUARD banana jack