Installing Tektronix socket adapters
The next graphic is a device test board. Notice that the test board contains the letters C, B, and E,
which also corresponds to the function of each terminal.
Figure 15: Model 8010-DTB-CT assembly
Measurement considerations
The following topics describe some considerations you should be aware of when taking
measurements.
Local and remote sense measurements
With the Model 8010, you can make measurements using either 2-wire (local) or 4-wire (remote)
sensing. Examples of each are shown in the "Model 8010 Interconnection Reference Guide."
The 2-wire (local) or 4-wire (remote) sensing is for DC IV measurements only. HV C-V
measurements made through the 8010 are 2-wire (local) sense only.