DMM4020
Technical Reference
Send the string “
EXIT
” to exit the calibration mode at any time. If a function has not
been completely calibrated, a CAL ERROR will be displayed.
Note
When the Meter returns a recommended calibration value followed by
“
W4
”, 4-wire compensation is enabled on the input.
Note
When in current calibration, if “Disconnect Test Leads!” is returned, all
test leads should be removed from the Meter.
Note
When “Connect to mA socket!” is returned from the Meter, use the mA
input terminal to connect to the calibrator. When “Connect to 10A socket”
is returned from the Meter, use the 10A input terminal to connect to the
calibrator.
Calibration Points
Listed below are the calibration values required for each calibration point of the
calibration procedure.
{//adc
{0.0, -0.0001999, -0.0001, 0.0001, 0.0001999}, range 1
{0.0, -0.001999, -0.001, 0.001, 0.001999}, range 2
{0.0, -0.01999, -0.01, 0.01, 0.01999}, range 3
{0.0, -0.1999, -0.1, 0.1, 0.1999}, range 4
{0.0, -1.999, -1.0, 1.0, 1.999}, range 5
{0.0, -10.0, -5.0, 5.0, 10.0}, range 6
},
{//aac
{0.0, 0.001, 0.005, 0.01, 0.01999}, range 1
{0.0, 0.01, 0.05, 0.1, 0.1999}, range 2
{0.0, 0.01, 0.5, 1.0, 1.999}, range 3
{0.0, 1.0, 2.0, 5.0, 10.0}, range 4
},
3-18
Содержание DMM4020
Страница 1: ...DMM4020 Digital Multimeter Technical Reference Revision B 077 0365 02...
Страница 6: ...DMM4020 Technical Reference iv...
Страница 8: ...DMM4020 Technical Reference vi...
Страница 10: ...DMM4020 Technical Reference 1 2...
Страница 24: ...DMM4020 Technical Reference 1 16...
Страница 26: ...DMM4020 Technical Reference 2 2...
Страница 34: ...DMM4020 Technical Reference 3 2...
Страница 52: ...DMM4020 Technical Reference 3 20...