Performance Verification and Calibration
Adjustment (Calibration)
3
Table 3-9. Adjustment Steps
Step
Func./Command
Point
Input Signal
Description
Volts DC: zero points using the 4-wire short
1
VDC
0 mV/0V
0 mV
Zero point of VDC, all ranges
Volts DC: Gains – Adjust using 5700A/5725A
2
VDC
-199.9 mV
-199.9 mV
Gain of VDC, 200 mV range
3
VDC
-100 mV
-100 mV
4
VDC
100 mV
100 mV
5
VDC
199.9 mV
199.9 mV
6
VDC
-1.999 V
-1.999 V
Gain of VDC, 2 V range
7
VDC
-1 V
-1 V
8
VDC
1 V
1 V
9
VDC
1.999 V
1.999 V
10
VDC
-19.99 V
-19.99 V
Gain of VDC, 20 V range
11
VDC
-10 V
-10 V
12
VDC
10 V
10 V
13
VDC
19.99 V
19.99 V
14
VDC
-199.9 V
-199.9 V
Gain of VDC, 200 V range
15
VDC
-100 V
-100 V
16
VDC
100 V
100 V
17
VDC
199.9 V
199.9 V
18
VDC
-1000 V
-1000 V
Gain of VDC, 1000 V range
19
VDC
-500 V
-500 V
20
VDC
500 V
500 V
21
VDC
1000 V
1000 V
Volts AC: Zero points – Adjust with 5700A in standby or disconnected
22
VAC
0 mV/0 V
0 mV
Zero point of VAC, all ranges
Volts AC: Gains – Adjust using 5700A/5725A
23
VAC
10 mV
10 mV@1 kHz
Gain of VAC, 200 mV range
24
VAC
100 mV
100 mV@1 kHz
25
VAC
150 mV
150 mV@1 kHz
26
VAC
199.9 mV
199.9 mV@1 kHz
27
VAC
0.1 V
0.1 V@1 kHz
Gain of VAC, 2 V range
28
VAC
1 V
1 V@1 kHz
29
VAC
1.5 V
1.5 V@1 kHz
30
VAC
1.999 V
1.999 V@1 kHz
3-13
Содержание DMM4020
Страница 1: ...DMM4020 Digital Multimeter Technical Reference Revision B 077 0365 02...
Страница 6: ...DMM4020 Technical Reference iv...
Страница 8: ...DMM4020 Technical Reference vi...
Страница 10: ...DMM4020 Technical Reference 1 2...
Страница 24: ...DMM4020 Technical Reference 1 16...
Страница 26: ...DMM4020 Technical Reference 2 2...
Страница 34: ...DMM4020 Technical Reference 3 2...
Страница 52: ...DMM4020 Technical Reference 3 20...