Performance Verification and Calibration
Adjustment (Calibration)
3
Table 3-9. Adjustment Steps (cont.)
Step
Func./Command
Point
Input Signal
Description
63
OHMS
10
Ω
10
Ω
, 4 Wire
Gain of OHMS, 20 k
Ω
range
64
OHMS
5 k
Ω
5k
Ω
, 4 Wire
65
OHMS
10 k
Ω
10 k
Ω
, 4 Wire
66
OHMS
15 k
Ω
15 k
Ω
, 4 Wire
67
OHMS
100
Ω
100
Ω
, 2 Wire
Gain of OHMS, 200 k
Ω
range
68
OHMS
50 k
Ω
50 k
Ω
, 2 Wire
69
OHMS
100 k
Ω
100 k
Ω
, 2 Wire
70
OHMS
150 k
Ω
150 k
Ω
, 2 Wire
71
OHMS
1 k
Ω
1 k
Ω
, 2 Wire
Gain of OHMS, 2 M
Ω
range
72
OHMS
0.5 M
Ω
0.5 M
Ω
, 2 Wire
73
OHMS
1 M
Ω
1 M
Ω
, 2 Wire
74
OHMS
1.5 M
Ω
1.5 M
Ω
, 2 Wire
75
OHMS
10 k
Ω
10 k
Ω
, 2 Wire
Gain of OHMS, 20 M
Ω
range
76
OHMS
5 M
Ω
5 M
Ω
, 2 Wire
77
OHMS
10 M
Ω
10 M
Ω
, 2 Wire
78
OHMS
15 M
Ω
15 M
Ω
, 2 Wire
79
OHMS
100 k
Ω
100 k
Ω
, 2 Wire
Gain of OHMS, 100 M
Ω
range
80
OHMS
20 M
Ω
20 M
Ω
, 2 Wire
81
OHMS
50 M
Ω
50 M
Ω
, 2 Wire
82
OHMS
100 M
Ω
100 M
Ω
, 2 Wire
Current AC: zero points – Adjust using 5700A/5725A
83
AAC
0 mA/0 A
Open, no connection Zero point of AAC, all ranges
Current AC: Gains – Adjust using 5700A/5725A
84
AAC
1 mA
1 mA@500 Hz
Gain of AAC, 20 mA range
85
AAC
5 mA
5 mA@500 Hz
86
AAC
10 mA
10 mA@500 Hz
87
AAC
19.99 mA
19.99 mA@500 Hz
88
AAC
10 mA
10 mA@500 Hz
Gain of AAC, 200 mA range
89
AAC
50 mA
50 mA@500 Hz
90
AAC
100 mA
100 mA@500 Hz
91
AAC
199.9 mA
199.9 mA@500 Hz
92
AAC
0.1 A
0.1 A@500 Hz
Gain of AAC, 2 A range
93
AAC
0.5 A
0.5 A@500 Hz
94
AAC
1 A
1 A@500 Hz
95
AAC
1.999 A
1.999 A@500 Hz
3-15
Содержание DMM4020
Страница 1: ...DMM4020 Digital Multimeter Technical Reference Revision B 077 0365 02...
Страница 6: ...DMM4020 Technical Reference iv...
Страница 8: ...DMM4020 Technical Reference vi...
Страница 10: ...DMM4020 Technical Reference 1 2...
Страница 24: ...DMM4020 Technical Reference 1 16...
Страница 26: ...DMM4020 Technical Reference 2 2...
Страница 34: ...DMM4020 Technical Reference 3 2...
Страница 52: ...DMM4020 Technical Reference 3 20...