DMM4020
Technical Reference
AC Voltage Specifications
AC Voltage specifications are for ac sinewave signals >5 % of range. For inputs from 1 % to 5 % of range and <50 kHz,
add an additional error of 0.1 % of range, and for 50 kHz to 100 kHz, add 0.13 % of range.
Maximum Input
................................................. 750 V rms or 1000 V peak or 8 x 10
7
Volts-Hertz product
Measurement Method
....................................... AC-coupled true-rms. Measures the ac component of input with up to
1000 V dc bias on any range.
AC Filter Bandwidth
......................................... 20 Hz – 100 kHz
Common Mode Rejection
................................. 60 dB at 50 Hz or 60 Hz (1 k
Ω
unbalance)
Maximum Crest Factor
.................................... 3:1 at Full Scale
Additional Crest Factor Errors (<100 Hz)
...... Crest Factor 1-2, 0.05 % of full scale
Crest Factor 2-3, 0.2 % of full scale
Only applies for non-sinusoid signals
Input Characteristics
Range
Full-Scale
(5-1/2 Digits)
Resolution
Input Impedance
Slow
Medium
Fast
200 mV
199.999 mV
1
µ
V
10
µ
V
10
µ
V
1 M
Ω
±
2 % shunted by
<100 pf
2 V
1.99999 V
10
µ
V
100
µ
V
100
µ
V
20 V
19.9999 V
100
µ
V
1000
µ
V
1000
µ
V
200 V
199.999 V
1 mV
10 mV
10 mV
750 V
750.00 V
10 mV
100 mV
100 mV
Accuracy
Range
Frequency
Uncertainty
[1]
Temperature
Coefficient/°C
Outside 18 – 28 °C
90 days
1 year
23 °C
±
5 °C
23 °C
±
5 °C
200 mV
20 Hz – 45 Hz
0.8 + 0.05
0.9 + 0.05
0.01 + 0.005
45 Hz – 20 kHz
0.15 + 0.05
0.2 + 0.05
0.01 + 0.005
20 kHz – 50 kHz
0.3 + 0.05
0.35 + 0.05
0.01 + 0.005
50 kHz – 100 kHz
0.8 + 0.05
0.9 + 0.05
0.05 + 0.01
2 V
20 Hz – 45 Hz
0.8 + 0.05
0.9 + 0.05
0.01 + 0.005
45 Hz – 20 kHz
0.15 + 0.05
0.2 + 0.05
0.01 + 0.005
20 kHz – 50 kHz
0.3 + 0.05
0.35 + 0.05
0.01 + 0.005
50 kHz – 100 kHz
0.8 + 0.05
0.9 + 0.05
0.05 + 0.01
20 V
20 Hz – 45 Hz
0.8 + 0.05
0.9 + 0.05
0.01 + 0.005
45 Hz – 20 kHz
0.15 + 0.05
0.2 + 0.05
0.01 + 0.005
20 kHz – 50 kHz
0.3 + 0.05
0.35 + 0.05
0.01 + 0.005
50 kHz – 100 kHz
0.8 + 0.05
0.9 + 0.05
0.05 + 0.01
200 V
20 Hz – 45 Hz
0.8 + 0.05
0.9 + 0.05
0.01 + 0.005
45 Hz – 20 kHz
0.15 + 0.05
0.2 + 0.05
0.01 + 0.005
20 kHz – 50 kHz
0.3 + 0.05
0.35 + 0.05
0.01 + 0.005
50 kHz – 100 kHz
0.8 + 0.05
0.9 + 0.05
0.05 + 0.01
750 V
20 Hz – 45 Hz
0.8 + 0.05
0.9 + 0.05
0.01 + 0.005
45 Hz – 20 kHz
0.15 + 0.05
0.2 + 0.05
0.01 + 0.005
20 kHz – 50 kHz
0.3 + 0.05
0.35 + 0.05
0.01 + 0.005
50 kHz – 100 kHz
0.8 + 0.05
0.9 + 0.05
0.05 + 0.01
Notes:
[1] Uncertainty given as
±
(% of r % of range)
1-12
Содержание DMM4020
Страница 1: ...DMM4020 Digital Multimeter Technical Reference Revision B 077 0365 02...
Страница 6: ...DMM4020 Technical Reference iv...
Страница 8: ...DMM4020 Technical Reference vi...
Страница 10: ...DMM4020 Technical Reference 1 2...
Страница 24: ...DMM4020 Technical Reference 1 16...
Страница 26: ...DMM4020 Technical Reference 2 2...
Страница 34: ...DMM4020 Technical Reference 3 2...
Страница 52: ...DMM4020 Technical Reference 3 20...