Performance Verification and Calibration
Introduction
3
Introduction
This chapter of the Calibration Manual provides performance tests to verify the Meter is
operating within published specifications as well as a complete calibration procedure.
The performance test and, if necessary, the calibration procedure can be performed both
periodically and after service or repair. A one-year calibration interval is recommended to
maintain the accuracy of the Meter.
The performance tests can be used as an acceptance test upon receipt of the Meter. Use
the 90-day specifications when performing an acceptance test after performing a
calibration.
Required Equipment
Table 3-1 lists the equipment required for performance testing and calibration of the
Meter.
Table 3-1. Required Test Equipment
Function
Instrument Type
Model
Comments
Volts DC
Calibrator
Fluke 5700A
Figure 3-1
4-wire short
Tektronix low thermal 4-
wire short or equivalent
Tektronix PN 013-0369-00
Volts AC
Calibrator
Fluke 5700A
Figure 3-1
Amplifier
Fluke 5725A
Resistance
Calibrator
Fluke 5700A
Figures 3-2 & 3-3
Amps DC
Calibrator
Fluke 5700A
Figure 3-4
Amplifier
Fluke 5725A
Figure 3-5
4-wire short
Tektronix low thermal 4-
wire short or equivalent
Tektronix PN 013-0369-00
Amps AC
Calibrator
Fluke 5700A
Figure 3-4
Amplifier
Fluke 5725A
Figure 3-5
Frequency
Function Generator
Fluke 271 or equivalent
Requires frequency specification
better than 0.0025 %
Resistance
Adjust
Calibrator
Fluke 5520A
Requires resistance accuracy better
than 0.0045 %
3-3
Содержание DMM4020
Страница 1: ...DMM4020 Digital Multimeter Technical Reference Revision B 077 0365 02...
Страница 6: ...DMM4020 Technical Reference iv...
Страница 8: ...DMM4020 Technical Reference vi...
Страница 10: ...DMM4020 Technical Reference 1 2...
Страница 24: ...DMM4020 Technical Reference 1 16...
Страница 26: ...DMM4020 Technical Reference 2 2...
Страница 34: ...DMM4020 Technical Reference 3 2...
Страница 52: ...DMM4020 Technical Reference 3 20...