General Maintenance
Power Considerations
2
Replacing the Fuses
The Meter uses one fuse to protect the line-power input and two fuses to protect the
current-measurement inputs.
Line-Power Fuse
The Meter has a line-power fuse in series with the power supply. Table 2-1 indicates the
proper fuse for each of the four line-voltage selections. The line-power fuse is accessed
through the rear panel.
1. Unplug the power cord.
2. Insert a small screwdriver blade into the narrow recess to the left of the fuse
holder and pry it to the right until the holder pops out. See Figure 2-1.
3. Remove the fuse and replace it with a fuse of an appropriate rating for the
selected line-power voltage. See Table 2-1.
4. Replace the selector block back into the fuse holder.
Warning
To avoid electric shock or fire, do not use makeshift fuses or
short-circuit the fuse holder.
Table 2-1. Line Voltage to Fuse Rating
Line Voltage Selection
Fuse Rating
100 / 120
0.
200
A, 250 V (slow blow)
220 / 240
0.
100
A, 250 V (slow blow)
eue20.eps
Figure 2-1. Replacing the Line Power Fuse
2-5