![Tektronix CSA7404B Скачать руководство пользователя страница 51](http://html1.mh-extra.com/html/tektronix/csa7404b/csa7404b_user-manual_1077969051.webp)
Incoming Inspection
CSA7000B Series & TDS7000B Series Instruments User Manual
1- 31
6.
Set the Vertical
SCALE
to
100 mV
per division.
7.
Set the time base:
Set the horizontal
SCALE
to
1 ms/div
. The time-base
readout is displayed at the bottom of the graticule.
8.
Save the settings:
a.
Pull down the
File
menu to select
Instrument Setup . . .
. This displays
the instrument setups control window.
b.
Touch the
Save
button under Save settings to file in the control window.
This displays a familiar Windows dialog box for choosing a destination
directory naming the file.
c.
In the Save Instrument Setup As dialog box, select the
3
1
/
2
Floppy (A:)
icon in the
Save in:
drop-down list to set the save destination to the
floppy disk.
d.
Note the default file name, and then touch the
Save
button to save the
setup to the default file name.
9.
Change the settings again:
Set the horizontal
SCALE
to
200
s/div
.
10.
Verify the file system works:
a.
Touch the
Recall Setups
tab in the control window.
b.
Touch the
Recall
button under Recall settings from file in the control
window. This displays a familiar Windows dialog box for locating the
settings file that you want to recall.
c.
In the Recall Instrument Setup dialog box, select the
3
1
/
2
Floppy (A:)
icon in the
Look in:
drop-down list.
d.
Locate and double-touch in the dialog box on the setup file that you
previously stored.
e.
Verify that the instrument retrieved the saved setup from the disk. Do
this by noticing the horizontal
SCALE
is again 1 ms and the waveform
shows ten cycles just as it did when you saved the setup.
11.
Remove the test hookup:
a.
Disconnect the equipment from the channel input and the probe
compensation output.
b.
Remove the floppy disk from the floppy disk drive.
Содержание CSA7404B
Страница 4: ......
Страница 16: ...Table of Contents xii CSA7000B Series TDS7000B Series Instruments User Manual ...
Страница 24: ...Product Description 1 4 CSA7000B Series TDS7000B Series Instruments User Manual ...
Страница 130: ...Acquiring Waveforms 3 62 CSA7000B Series TDS7000B Series Instruments User Manual ...
Страница 186: ...Triggering 3 118 CSA7000B Series TDS7000B Series Instruments User Manual ...
Страница 234: ...Measuring Waveforms 3 166 CSA7000B Series TDS7000B Series Instruments User Manual ...
Страница 288: ...Creating and Using Math Waveforms 3 220 CSA7000B Series TDS7000B Series Instruments User Manual ...
Страница 324: ...Data Input Output 3 256 CSA7000B Series TDS7000B Series Instruments User Manual ...
Страница 330: ...MyScope 3 262 CSA7000B Series TDS7000B Series Instruments User Manual ...
Страница 364: ...Appendix A Specifications A 30 CSA7000B Series TDS7000B Series Instruments User Manual ...
Страница 378: ...Appendix B Automatic Measurements Supported B 14 CSA7000B Series TDS7000B Series Instruments User Manual ...
Страница 414: ...Glossary Glossary 18 CSA7000B Series TDS7000B Series Instruments User Manual ...
Страница 437: ...Index CSA7000B Series TDS7000B Series Instruments User Manual Index 23 setup 3 122 C 5 C 6 Size C 5 ...
Страница 438: ...Index Index 24 CSA7000B Series TDS7000B Series Instruments User Manual ...