Appendix B: Automatic Measurements Supported
CSA7000B Series & TDS7000B Series Instruments User Manual
B- 3
Table B- 1: Supported measurements and their definition (Cont.)
Name
Definition
Positive Overshoot
Voltage measurement over the entire waveform or gated region.
PositiveOvershoot
=
Max–High
Amplitude
×
100%
Positive Width
Timing measurement of the first pulse in the waveform or gated region. The distance (time)
between MidRef (default 50%) amplitude points of a positive pulse.
Rise time
Timing measurement. Time taken for the leading edge of the first pulse in the waveform or
gated region to rise from a Low Ref value (default = 10%) to a High Ref value (default = 90%) of
its final value.
RMS
Voltage measurement. The true Root Mean Square voltage over the entire waveform or gated
region.
Mean
The average of all acquired points within (or on) the histogram box.
Median
Half of all acquired points within (or on) the histogram box are less than and half are greater
than this value.
StdDev
The standard deviation (Root Mean Square (RMS) deviation) of all acquired points within (or on)
the histogram box.
Hits in Box
Displays the number of points in the histogram box or on the box boundary.
Waveform Count
Displays the number of waveforms that have contributed to the histogram.
Peak Hits
Displays the number of points in the largest bin of the histogram.
Pk-Pk
Displays the peak-to-peak value of the histogram. Vertical histograms display the “voltage” of
the highest nonzero bin minus the “voltage” of the lowest nonzero bin. Horizontal histograms
display the “time” of the rightmost nonzero bin minus the “time” of the leftmost nonzero bin
(Max -- Min).
Max
Displays the maximum voltage or time.
Min
Displays the minimum voltage or time.
Mean
1 StdDev
The percentage of points in the histogram that are within 1 standard deviation of the histogram
mean.
Mean
2 StdDev
The percentage of points in the histogram that are within 2 standard deviations of the histogram
mean.
Mean
3 StdDev
The percentage of points in the histogram that are within 3 standard deviations of the histogram
mean.
Содержание CSA7404B
Страница 4: ......
Страница 16: ...Table of Contents xii CSA7000B Series TDS7000B Series Instruments User Manual ...
Страница 24: ...Product Description 1 4 CSA7000B Series TDS7000B Series Instruments User Manual ...
Страница 130: ...Acquiring Waveforms 3 62 CSA7000B Series TDS7000B Series Instruments User Manual ...
Страница 186: ...Triggering 3 118 CSA7000B Series TDS7000B Series Instruments User Manual ...
Страница 234: ...Measuring Waveforms 3 166 CSA7000B Series TDS7000B Series Instruments User Manual ...
Страница 288: ...Creating and Using Math Waveforms 3 220 CSA7000B Series TDS7000B Series Instruments User Manual ...
Страница 324: ...Data Input Output 3 256 CSA7000B Series TDS7000B Series Instruments User Manual ...
Страница 330: ...MyScope 3 262 CSA7000B Series TDS7000B Series Instruments User Manual ...
Страница 364: ...Appendix A Specifications A 30 CSA7000B Series TDS7000B Series Instruments User Manual ...
Страница 378: ...Appendix B Automatic Measurements Supported B 14 CSA7000B Series TDS7000B Series Instruments User Manual ...
Страница 414: ...Glossary Glossary 18 CSA7000B Series TDS7000B Series Instruments User Manual ...
Страница 437: ...Index CSA7000B Series TDS7000B Series Instruments User Manual Index 23 setup 3 122 C 5 C 6 Size C 5 ...
Страница 438: ...Index Index 24 CSA7000B Series TDS7000B Series Instruments User Manual ...