![Tektronix CSA7404B Скачать руководство пользователя страница 101](http://html1.mh-extra.com/html/tektronix/csa7404b/csa7404b_user-manual_1077969101.webp)
Acquiring Waveforms
CSA7000B Series & TDS7000B Series Instruments User Manual
3- 33
Use the procedure that follows to set up roll mode acquisitions.
Overview
To set Roll Mode
Control elements and resources
Prerequisites
1.
The horizontal and vertical controls must be set up.
Triggering should also be set up.
See page 3--29 for acquisition setup and
page 3--63 for trigger setup.
To enable roll
mode
2.
Touch the Horiz button. Select the Acquisition tab from
the Horiz/Acq control window, or select Horizontal/Ac-
quisition Setup from the Horiz/Acq menu to display the
Acquisition Mode control window. Select the Acquisition
tab.
3.
Select Roll Mode AUTO to enable roll mode.
When the horizontal scale is 100 ms per division at a
record length of 500 points and the acquisition mode is
Sample or Pk Detect, roll mode turns on. As the record
length becomes larger, the time per division required to
enter roll mode becomes slower.
Note.
Envelope, Average, FastAcq, and Waveform
Database acquisition modes inhibit roll mode.
To single
sequence roll
mode
4.
Push the SINGLE button to start acquiring and acquire
enough waveforms to satisfy the acquisition mode and
then stop.
To turn off roll
mode
acquisitions
5.
Do the following step to stop acquisitions in roll mode:
H
If you are not in Single Sequence, push RUN/
STOP to stop roll mode.
H
If you are in Single Sequence, roll mode acquisi-
tions stop automatically when a complete record is
acquired.
To Set Roll Mode
Содержание CSA7404B
Страница 4: ......
Страница 16: ...Table of Contents xii CSA7000B Series TDS7000B Series Instruments User Manual ...
Страница 24: ...Product Description 1 4 CSA7000B Series TDS7000B Series Instruments User Manual ...
Страница 130: ...Acquiring Waveforms 3 62 CSA7000B Series TDS7000B Series Instruments User Manual ...
Страница 186: ...Triggering 3 118 CSA7000B Series TDS7000B Series Instruments User Manual ...
Страница 234: ...Measuring Waveforms 3 166 CSA7000B Series TDS7000B Series Instruments User Manual ...
Страница 288: ...Creating and Using Math Waveforms 3 220 CSA7000B Series TDS7000B Series Instruments User Manual ...
Страница 324: ...Data Input Output 3 256 CSA7000B Series TDS7000B Series Instruments User Manual ...
Страница 330: ...MyScope 3 262 CSA7000B Series TDS7000B Series Instruments User Manual ...
Страница 364: ...Appendix A Specifications A 30 CSA7000B Series TDS7000B Series Instruments User Manual ...
Страница 378: ...Appendix B Automatic Measurements Supported B 14 CSA7000B Series TDS7000B Series Instruments User Manual ...
Страница 414: ...Glossary Glossary 18 CSA7000B Series TDS7000B Series Instruments User Manual ...
Страница 437: ...Index CSA7000B Series TDS7000B Series Instruments User Manual Index 23 setup 3 122 C 5 C 6 Size C 5 ...
Страница 438: ...Index Index 24 CSA7000B Series TDS7000B Series Instruments User Manual ...