Measuring Waveforms
3- 148
CSA7000B Series & TDS7000B Series Instruments User Manual
Overview
Related control elements and resources
To take automatic measurements (Cont.)
Set
measurement
reference
levels (Cont.)
18.
To select how the instrument determines the base and
top of the waveform, touch Min-Max, Histogram, or
Histogram mean.
H
Min-max. Uses the highest and lowest values of the
waveform record. This setting is best for examining
waveforms that have no large, flat portions at a
common value, such as sine waves and triangle
waves — almost any waveform except for pulses.
See Figure 3--36 on page 3--143.
H
Histogram. Selects the most common values above
or below the midpoint. Since this approach ignores
short term aberrations (overshoot, ringing, and so
on), Histogram is the best setting for examining
pulses. See Figure 3--36 on page 3--143.
H
Histogram mean. Calculates the mean value using
all values either above or below the midpoint
(depending on whether it is defining the high or low
reference level). Histogram mean is best for
examining eye patterns and optical signals.
19.
To select the reference level units, touch Units Absolute
or Percentage.
H
Absolute. Sets the units to absolute values in user
units.
H
Percentage. Sets the units as a percentage of the
High/Low range.
20.
To set the reference levels, touch HighRef, Mid Ref, Low
Ref, or Mid2 Ref, and use the multipurpose knobs or
keypad to set the levels.
Содержание CSA7404B
Страница 4: ......
Страница 16: ...Table of Contents xii CSA7000B Series TDS7000B Series Instruments User Manual ...
Страница 24: ...Product Description 1 4 CSA7000B Series TDS7000B Series Instruments User Manual ...
Страница 130: ...Acquiring Waveforms 3 62 CSA7000B Series TDS7000B Series Instruments User Manual ...
Страница 186: ...Triggering 3 118 CSA7000B Series TDS7000B Series Instruments User Manual ...
Страница 234: ...Measuring Waveforms 3 166 CSA7000B Series TDS7000B Series Instruments User Manual ...
Страница 288: ...Creating and Using Math Waveforms 3 220 CSA7000B Series TDS7000B Series Instruments User Manual ...
Страница 324: ...Data Input Output 3 256 CSA7000B Series TDS7000B Series Instruments User Manual ...
Страница 330: ...MyScope 3 262 CSA7000B Series TDS7000B Series Instruments User Manual ...
Страница 364: ...Appendix A Specifications A 30 CSA7000B Series TDS7000B Series Instruments User Manual ...
Страница 378: ...Appendix B Automatic Measurements Supported B 14 CSA7000B Series TDS7000B Series Instruments User Manual ...
Страница 414: ...Glossary Glossary 18 CSA7000B Series TDS7000B Series Instruments User Manual ...
Страница 437: ...Index CSA7000B Series TDS7000B Series Instruments User Manual Index 23 setup 3 122 C 5 C 6 Size C 5 ...
Страница 438: ...Index Index 24 CSA7000B Series TDS7000B Series Instruments User Manual ...