Performance Check/Calibration—Type 324
Preliminary Procedure for Performance Check Only
1.
Connect the Type 324 to the output of the vari
able DC power supply (set for +8 volts) with the two
patch cords. If step 8 is deleted, the instrument can be
connected directly to an AC power source which meets
the voltage and frequency requirements of this instru
ment.
NOTE
Battery operation can be used for this procedure
if the internal batteries are fully charged before
proceeding.
2.
Set the controls as given under Preliminary Con
trol Settings. Allow at least five minutes warmup before
proceeding.
Preliminary Control Settings
Set the Type 324 controls as follows:
Vertical Controls
IN PU T
GND
Vertical POSITION
Midrange
Triggering Controls
T R IG G E R
+ A U T 0
A TTE N (side panel)
3. Begin the performance check with step 8.
Preliminary Procedure for Complete Calibration
1.
Remove the cabinet from the Type 324 (see
Section 2).
Horizontal Controls
T IM E /D IV
V A R IA B L E
Horizontal POSITION
X 5 H O R IZ MAG
1 ms
CAL
Midrange
Pushed in
2.
Remove the high-voltage cover and power pack
from the Type 324. Removal instructions are given in
CRT Controls
the Maintenance section.
FOCUS
3.
Using the AC power cord, connect the power pack
of the Type 324 to an AC line voltage source which is
within the voltage and frequency requirements of this
instrument.
4.
Set the controls as given under Preliminary Con
trol Settings. Allow at least five minutes warmup before
proceeding.
IN T E N S IT Y
Power Controls
POWER
Power Pack
(rear panel)
Adjust for well
defined display
Midrange
ON
F U L L CHG
NOTES
5-8