Performance Check/Calibration—Type 324
R E Q U IR E M E N T: Stable display in IN T T R IG AC
position of the Trig/Horiz Coupling switch with the
T R IG G E R control set to +A U TO , variable positive-
slope area, variable negative-slope area, and —A U TO ;
check with a 0.3-division display at 30 hertz.
PERFORM ANCE: All c o rre c t_____ ; incorrect (list
exceptions)________________________________.
43. Check Low-Frequency Reject Operation
page 5-28
R E Q U IR EM EN T: Stable display with 0.3-division dis
play at 15 kilohertz with the T R IG G E R control set to
+AUTO , variable positive-slope area, variable negative-
slope area, and —A U TO . Check that stable display can
not be obtained at 30 hertz with any setting of the
TR IG G ER control.
PERFORMANCE: Correct
; incorrect
44. Check Trigger Slope Operation
page 5-29
REQ U IR EM EN T: Stable triggering on correct slope of
trigger signal in +A U TO , variable positive-slope area,
variable negative-slope area, and —AU TO positions of
the TR IG G ER control.
PERFORMANCE: All c o r r e c t______; incorrect (list
exceptions)
45. Check Trigger Control Range
page 5-29
R EQ U IR EM EN T: A t least + and - 0 . 8 volts with E X T
T R IG A TTE N switch set to 1X. Check in variable
positive-slope area and variable negative-slope area.
PERFORMANCE: Correct
; incorrect
R EQ U IR EM EN T: A t least + and —8 volts with EXT
T R IG A TTEN switch set to 10X. Check in variable
positive-slope area and variable negative-slope area.
PERFORM ANCE: Correct
; incorrect
46. Check Calibrator Operation
page 5-30
R EQ U IR EM EN T: Two microseconds or less risetime.
PERFORM ANCE:
microseconds.
R E Q U IR EM EN T: Duration of one cycle between 5
and 8.3 divisons at .2 m s/D IV .
PERFORM ANCE:
divisions.
R E Q U IR E M E N T : Length of positive segment of
square wave between 4.0 and 6.0 divisions with one
complete cycle/10 divisions.
PERFO RM ANCE:
divisions.
R E Q U IR E M E N T : Correct amplitude at CAL O U T
jack and in the 5 D IV CAL position of the V O L T S /
D IV switch.
PERFO RM ANCE: C o r r e c t ______; in c o rre c t______ .
COMPLETE PERFORMANCE
C H EC K /C A LIB R A TIO N PROCEDURE
General
The following procedure is arranged so the Type 324
can be calibrated with the least interaction of adjust
ments and reconnection of equipment. A picture of the
test equipment required for each group of steps is given
to aid in identification of the necessary equipment. The
control settings and test equipment setup throughout
th is procedure continue from the preceding step(s)
unless noted otherwise. The control settings can be
checked at any test equipment required picture by set
ting the controls as given at the start of the procedure
under the heading Preliminary Control Settings. Then
make any changes listed beneath the test equipment
required picture under the heading Partial Procedure
(also applies to partial calibration procedure).
NOTE
Control titles which are printed on the outside of
the Type 324 are capitalized in this procedure (e.g.,
VOLTS/D/V). Internal adjustments are initial capi
talized only (e.g.. Limit Centering).
The following procedure uses the equipment listed
under Test Equipment Required. If other equipment is
substituted, control settings or calibration setup may
need to be altered to meet the requirements of the
equipment used. Detailed operating instructions for the
test equipment are not given in this procedure. Refer to
the instruction manual for the test equipment if more
information is required.
NOTE
This instrument should be calibrated at an ambient
temperature of +25°C ±5°C for best overall accur
acy. The performance of this instrument can be
checked at any temperature within the 0 °C to
+40°C range. If the ambient temperature is outside
the given range, see Section 1 for the applicable
tolerances.