Performance Check/Cafibration—Type 324
ag. Turn the T R IG G E R control counterclockwise to
the variable negative-slope area.
ah. CHECK—Stable CRT display can be obtained
with the Trig/Horiz Coupling switch set to EXT T R IG
AC and DC (TR IG G E R control may be adjusted as
necessary to obtain a stable display).
ai. Turn the T R IG G E R control counterclockwise to
the variable positive-slope area.
aj. CHECK—Stable CRT display can be obtained with
the Trig/Horiz Coupling switch set to E X T T R IG AC
and DC (TR IG G E R control may be adjusted as neces
sary to obtain a stable display).
ak. Set the T R IG G E R control to +A U TO .
al. CHECK—Stable C RT display is presented,
am. Disconnect the high-frequency generator.
42. Check Low-Frequency Triggering Operation
a. Connect the low-frequency constant-amplitude sine-
wave generator to the V E R T INPUT connector through
the 42-inch 50-ohm BNC cable, 50-ohm BNC termin
ation and the BNC T connector. Connect the output
of the BNC T connector to the E X T T R IG OR H O R IZ
IN PU T connector with an 18-inch 50-ohm BNC cable.
b. Change the following control settings:
T IM E /D IV
10 ms
X 5 H O R IZ MAG
Pushed in
c.
Set the low-frequency generator for a 0.9-division
display (90 millivolts) at 30 hertz.
d.
CHECK—Stable CRT display is presented with the
Trig/Horiz Coupling switch set to E X T T R IG AC and
DC.
e.
Turn the T R IG G E R control clockwise to the vari
able positive-slope area.
f.
CHECK—Stable CRT display can be obtained with
the Trig/Horiz Coupling switch set to E X T T R IG AC
and DC (TR IG G E R control may be adjusted as neces
sary to obtain a stable display).
g. Turn the T R IG G E R control clockwise to the vari
able negative-slope area.
h. CHECK—Stable C RT display can be obtained with
the Trig/Horiz Coupling switch set to E X T T R IG AC
and DC (TR IG G E R Control may be adjusted as neces
sary to obtain a stable display).
i. Set the TR IG G E R control to —A U T O .
j. CHECK—Stable CRT display is presented with the
Trig/Horiz Coupling Switch set to E X T T R IG AC and
DC.
k. Set the Trig/Horiz Coupling switch to IN T T R IG
AC.
I. Set the low-frequency generator for a 0.3-division
display at 30 hertz.
m. C H EC K —Stable CRT display is presented.
n. Turn the T R IG G E R control counterclockwise to
the variable negative-slope area.
o. CHECK—Stable C RT display can be obtained with
the T R IG G E R control.
p. Turn the T R IG G E R control counterclockwise to
the variable positive-slope area.
q. CHECK—Stable CRT display can be obtained with
the TR IG G E R control.
r. Set the T R IG G E R control to +A U TO .
s. CHECK—Stable CRT display is presented.
43. Check Low-Frequency Reject Operation
a. Change the following control settings:
Trig/Horiz Coupling
IN T T R IG ACLF
REJ
T IM E /D IV
.1 ms
b. Set the low-frequency generator for a 0.3-division
display at 15 kilohertz.
5-28