![Tektronix 2213 Скачать руководство пользователя страница 59](http://html.mh-extra.com/html/tektronix/2213/2213_instruction-manual_1078827059.webp)
Performance Check Procedure—2213 Service
HORIZONTAL
Equipment Required (see Table 4-1):
Calibration Generator (Item 1)
Two 50-S2 BNC Cables (Item 4)
Leveled Sine-Wave Generator (Item 2)
Two 50-S2 BNC Terminations (Item 5)
Time-Mark Generator (Item 3)
INITIAL CONTROL SETTINGS
PROCEDURE STEPS
POWER
ON (button in)
CRT
AUTO INTENSITY
As desired
AUTO FOCUS
Best focused display
1. Check Timing Accuracy
a.
Connect 50-ns time markers from the time-mark
generator via a 5 0 -fi cable and a 5 0 -fi term ination to the
CH 1 OR X input connector. Connect the generator Trigger
output via a 50-£2 cable and a 50-f2 term ination to the
EXT INPUT connector.
Vertical
Channel 1 POSITION
V E R TIC A L MODE
CH 1 VO LTS/D IV
CH 1 VO LTS/DIV
Variable
INVERT
Channel 1 AC-GND-DC
Channel 2 AC-GND-DC
Midrange
CH 1
0.5 V
CAL detent
Normal (button out)
DC
GND
b.
Use the Channel 1 POSITION control to center the
trace vertically. Adjust the TRIGGER LEVEL control for
a stable, triggered display.
c.
Use the Horizontal POSITION control to align the
2nd tim e marker w ith the 2nd vertical graticule line.
NOTE
When making tim ing measurements, use as a reference
the same p o in t on each tim e marker.
Horizontal
POSITION
H O R IZO N TAL MODE
SEC/DIV
SEC/DIV Variable
X10 Magnifier
D ELAY TIME
M ULTIPLIER
Midrange
NO D LY
0.05 jus
C AL detent
O ff (knob in)
0.5 jus
<X 1
d.
CHECK—Timing accuracy is w ithin 3% (0.24 division)
at the 10th vertical graticule line. Exclude the firs t 50 ns
o f the sweep from the measurement.
e.
Repeat parts b through d fo r the remaining SEC/DIV
switch settings and corresponding time-mark generator
output (Normal) settings given in Table 4-4.
Trigger
VAR HOLDOFF
MODE
SLOPE
LEVEL
I NT
SOURCE
EXT COUPLING
NORM
NORM
S
Midrange
VERT MODE
EXT
DC-r10
f. Pull out the X10 Magnifier knob.
g.
CHECK—Magnified tim ing accuracy fo r each SEC/
D IV switch setting and corresponding time-mark generator
output (X10 Magnified) setting given in Table 4-4. A t each
setting combination, tim ing must be accurate w ithin 5%
(0.4 division) at the 10th vertical graticule line. Exclude
the first
50
ns o f the
sweep
and any
portion past the 100th
magnified division.
4-6
Содержание 2213
Страница 1: ...Ttektronix 2213 OSCILLOSCOPE SERVICE IN S T R U C T IO N M A N U A L ...
Страница 9: ...2213 Service 3827 01 viii The 2213 Oscilloscope ...
Страница 30: ...Theory of Operation 2213 Service 3 3 Figure 3 1 Basic block diagram of the 2213 Oscilloscope ...
Страница 67: ...Adjustment Procedure 2213 Service 5 2 ...
Страница 139: ...Figure 9 4 2213 block diagram S105 ...
Страница 140: ...a o m 7 x C O o z CHANNEL SWITCH AND VERTICAL OUTPUT ...
Страница 141: ...2213 Service ...
Страница 144: ...2213 ...
Страница 146: ...2213 3827 20 R V FB 4982 C 4 ...
Страница 147: ...S90I 4 5 ...
Страница 148: ...CIRCUIT BOARD INTERCONNECTIONS K tv ra n m CIRCUIT BO ARD IN T E R C O N N E C T IO N S ...
Страница 151: ......
Страница 153: ...2213 3 K T i RtV D E C ...
Страница 154: ......
Страница 155: ...2213 Service 1 3 a 7 ...
Страница 161: ...2213 Service n 3 a ...
Страница 168: ...s RE 2213 ...
Страница 169: ... CH 1 CH 2 VERTICAL PREAMPS ...
Страница 175: ......
Страница 177: ...TRIGGER ...
Страница 180: ...fr ...
Страница 181: ...f E V AUG I98Z SWEEP GENERATOR LOGIC ...
Страница 184: ......
Страница 188: ...2 2 1 3 lAlO FWTW m ain b o a rd ...
Страница 190: ...i _____H M N XY AMPLIFIER HORIZ OUTPUT ...
Страница 191: ...A 1 1 C 1 a i F 1 G 1 H 1 3 4 5 6 7 HO COM TO P 4 C C O 5 S I i S2T 2e R EV Dec I9BI 2 2 3 ...
Страница 203: ...2 2 1 3 Service R887 ASTIG R870 GEOM C784 5ns TIMING R860 GRID BIAS ...
Страница 207: ... ...
Страница 216: ...I ...
Страница 217: ...2213 OSCILLOSCOPE ...
Страница 218: ......
Страница 219: ...2213 OSrs ...
Страница 225: ......