Preventive Maintenance
7–2
1735 HD High Definition Waveform Monitor
A 2% RMA flux content solder is recommended for making repairs in this
instrument. Cleaning of rosin residue is not recommended. Most cleaning
solvents tend to reactivate the rosin and spread it under components where it may
cause corrosion under humid conditions. The rosin residue, if left alone, does not
exhibit these corrosive properties.
If the circuit board assemblies must be removed for cleaning, follow the
instructions for removal/ replacement under the heading of Corrective Mainte-
nance.
After cleaning, allow the interior to dry thoroughly before applying power to the
instrument.
Visual Inspection
After cleaning, check the instrument carefully for improperly seated transistors
or integrated circuits, defective connections, and damaged parts. To prevent
additional damage in the case of heat-damaged parts, determine the cause of
overheating before replacing the damaged part.
Periodic checks of the transistors and integrated circuits are not recommended.
The best measure of performance is the actual operation of the component in the
circuit.
Static-sensitive Components
This instrument contains electrical components that are susceptible to damage
from static discharge. Static voltages from 1 kV to 30 kV are common in
unprotected environments. Table 7–1 shows the relative static discharge
susceptibility of various semiconductor classes.
Table 7–1: Static susceptibility
Relative susceptibility levels
Voltage*
1
MOS and CMOS
100 to 500 V
2
ECL
200 to 500 V
3
SCHOTTKY SIGNAL DIODES
250 V
4
SCHOTTKY TTL
500 V
5
HF BIPOLAR TRANSISTORS
400 to 600 V
6
JFETS
600 to 800 V
7
LINEAR CIRCUITS
400 to 1000 V est.
8
LOW POWER SCHOTTKY TTL
900 V
Содержание 070-8462-03
Страница 5: ......
Страница 13: ...Table of Contents viii 1735 HD High Definition Waveform Monitor...
Страница 19: ...Preface xiv 1735 HD High Definition Waveform Monitor...
Страница 20: ...Getting Started...
Страница 21: ......
Страница 34: ...Installation 1735 HD High Definition Waveform Monitor 1 13 Figure 1 9 Typical custom installation...
Страница 35: ...Installation 1 14 1735 HD High Definition Waveform Monitor...
Страница 36: ...Operating Basics...
Страница 37: ......
Страница 63: ...Operating Basics 2 26 1735 HD High Definition Waveform Monitor...
Страница 64: ...Specifications...
Страница 65: ......
Страница 75: ...Specifications 3 10 1735 HD High Definition Waveform Monitor...
Страница 77: ......
Страница 78: ...Theory of Operation...
Страница 79: ......
Страница 96: ...Theory of Operation 1735 HD High Definition Waveform Monitor 4 17 Figure 4 2 1735 HD Clamp pulse generator...
Страница 109: ...Theory of Operation 4 30 1735 HD High Definition Waveform Monitor...
Страница 110: ...Performance Verification...
Страница 111: ......
Страница 134: ...Adjustment Procedure...
Страница 135: ......
Страница 158: ...Maintenance...
Страница 159: ......
Страница 163: ...Preventive Maintenance 7 4 1735 HD High Definition Waveform Monitor...
Страница 168: ...Troubleshooting 1735 HD High Definition Waveform Monitor 7 9 Figure 7 3 Multiple pin connectors...
Страница 179: ...Corrective Maintenance 7 20 1735 HD High Definition Waveform Monitor...
Страница 181: ...Repackaging 7 22 1735 HD High Definition Waveform Monitor Figure 7 10 Repackaging...
Страница 182: ...Replaceable Electrical Parts...
Страница 183: ......
Страница 252: ...Diagrams...
Страница 253: ......
Страница 256: ...Replaceable Mechanical Parts...
Страница 257: ......
Страница 265: ...Replaceable Mechanical Parts 10 8 1735 HD High Definition Waveform Monitor...
Страница 266: ...Appendices...
Страница 267: ......
Страница 270: ......
Страница 271: ......