Performance Verification
1735 HD High Definition Waveform Monitor
5–17
m. Set the DVM range for three-place accuracy and adjust the peak-to-peak
de level for a reading as close to zero as possible. Note the
reference reading.
n. Use the leveled sine-wave generator Frequency controls to provide all
frequencies from 250 kHz to 30 MHz.
o. Check frequency response for
±
2% or 14 mV from 250 kHz to 30 MHz.
p. Repeat steps j. through o. for CH 2B, CH1B, CH 3A, CH 2A, and
CH 1A.
q. Return the 1735 HD to X1 Gain.
17. Check PIX MON OUT Frequency Response
Requirement: Frequency Response within
±
5% of 50 kHz, up to 30 MHz.
a. Connect the peak-to-peak detector head to PIX MON OUT.
b. Connect the leveled sine-wave generator to the 1735 HD CH1 A.
Terminate the loop-through in 75
W
.
c.
Select CH 1A INPUT on the 1735 HD.
d. Set the leveled sine-wave generator Frequency Range to 50 kHz and
adjust its Output Amplitude for a 700 mV display on the 1735 HD.
e.
Set the DVM range for three-place accuracy and adjust the peak-to-peak
detector level for a reading as close to zero as possible. Note the
reference reading.
f.
Use the leveled sine-wave generator Frequency controls to provide all
frequencies from 250 kHz to 30 MHz.
g. Check that display voltage on the DVM is equal to that noted at 50 kHz,
±
5% (35 mV).
h. Disconnect all inputs and 75
W
terminations from the 1735 HD.
18. Check CH 1B Subtract
Requirement: A INPUT: >30 dB (<22 mV) from 50 kHz to 15 MHz. >23 dB
(<49.5 mV) from 15 MHz to 30 MHz.
a. Connect the output cable from the leveled sine wave generator, through a
50
W
precision cable, a 50-to-75
W
minimum loss attenuator, a 75
W
in-line termination and dual-input connector, to the 1735 HD CH1A and
CH 1B INPUTs.
Содержание 070-8462-03
Страница 5: ......
Страница 13: ...Table of Contents viii 1735 HD High Definition Waveform Monitor...
Страница 19: ...Preface xiv 1735 HD High Definition Waveform Monitor...
Страница 20: ...Getting Started...
Страница 21: ......
Страница 34: ...Installation 1735 HD High Definition Waveform Monitor 1 13 Figure 1 9 Typical custom installation...
Страница 35: ...Installation 1 14 1735 HD High Definition Waveform Monitor...
Страница 36: ...Operating Basics...
Страница 37: ......
Страница 63: ...Operating Basics 2 26 1735 HD High Definition Waveform Monitor...
Страница 64: ...Specifications...
Страница 65: ......
Страница 75: ...Specifications 3 10 1735 HD High Definition Waveform Monitor...
Страница 77: ......
Страница 78: ...Theory of Operation...
Страница 79: ......
Страница 96: ...Theory of Operation 1735 HD High Definition Waveform Monitor 4 17 Figure 4 2 1735 HD Clamp pulse generator...
Страница 109: ...Theory of Operation 4 30 1735 HD High Definition Waveform Monitor...
Страница 110: ...Performance Verification...
Страница 111: ......
Страница 134: ...Adjustment Procedure...
Страница 135: ......
Страница 158: ...Maintenance...
Страница 159: ......
Страница 163: ...Preventive Maintenance 7 4 1735 HD High Definition Waveform Monitor...
Страница 168: ...Troubleshooting 1735 HD High Definition Waveform Monitor 7 9 Figure 7 3 Multiple pin connectors...
Страница 179: ...Corrective Maintenance 7 20 1735 HD High Definition Waveform Monitor...
Страница 181: ...Repackaging 7 22 1735 HD High Definition Waveform Monitor Figure 7 10 Repackaging...
Страница 182: ...Replaceable Electrical Parts...
Страница 183: ......
Страница 252: ...Diagrams...
Страница 253: ......
Страница 256: ...Replaceable Mechanical Parts...
Страница 257: ......
Страница 265: ...Replaceable Mechanical Parts 10 8 1735 HD High Definition Waveform Monitor...
Страница 266: ...Appendices...
Страница 267: ......
Страница 270: ......
Страница 271: ......