Performance Verification
5–16
1735 HD High Definition Waveform Monitor
16. Check X1 and X5 Flat Response
Requirement: Flat response with 50kHz as a reference; 50 kHz to 30 MHz
within
±
2%. X5 Flat response with 50 kHz as a reference; 250 kHz to
30 MHz, within
±
2%.
a. Connect the leveled sine-wave generator Output, through a 50
W
precision cable and a 50-to-75
W
minimum loss attenuator, to the
1735 HD CH1A INPUT.
b. Connect the peak-to-peak detector head to the remaining side of the
1735 HD CH 1A INPUT. Connect the peak-to-peak detector head to the
+ (plus) input of the peak-to-peak detector. Connect the output of the
peak-to-peak detector, through a bnc-to-dual banana plug adapter, to the
Digital Voltmeter (DVM). Enable the peak-to-peak de (plus)
input.
c.
Select CH 1A INPUT, FLAT Filter, and ONE LINE Sweep on the
1735 HD.
d. Set the leveled sine-wave generator Frequency Range to 50 kHz and
adjust its Output Amplitude for a 700 mV display on the 1735 HD.
e.
Set the DVM range for three-place accuracy and adjust the peak-to-peak
de level for a reading as close to zero as possible. Note the
reference reading.
NOTE
. To calculate the deviation from flatness, set the leveled sine-wave
generator to the frequency being checked and adjust the output amplitude to
obtain the same level on screen as seen at the 50 kHz reference. Note the DVM
reading at the frequency being checked and subtract it from the DVM reading
observed at the 50 kHz reference.
f.
Use the leveled sine-wave generator Frequency controls to provide all
frequencies from 250 kHz to 30 MHz.
g. Check frequency response for
±
2% or 14 mV from 250 kHz to 30 MHz.
h. Repeat steps a. through g. for CH 2A, CH3A, CH 1B, CH 2B, and
CH 3B.
i.
Select the 1735 HD X5 Gain.
j.
Connect the leveled sine-wave generator Output to the 1735 HD CH 3B
INPUT. Select CH 3B INPUT on the 1735 HD.
k. Connect the peak-to-peak detector Head to the CH 3B INPUT.
l.
Set the leveled sine-wave generator to 50 kHz and adjust its Output
Amplitude for a 700 mV display.
Содержание 070-8462-03
Страница 5: ......
Страница 13: ...Table of Contents viii 1735 HD High Definition Waveform Monitor...
Страница 19: ...Preface xiv 1735 HD High Definition Waveform Monitor...
Страница 20: ...Getting Started...
Страница 21: ......
Страница 34: ...Installation 1735 HD High Definition Waveform Monitor 1 13 Figure 1 9 Typical custom installation...
Страница 35: ...Installation 1 14 1735 HD High Definition Waveform Monitor...
Страница 36: ...Operating Basics...
Страница 37: ......
Страница 63: ...Operating Basics 2 26 1735 HD High Definition Waveform Monitor...
Страница 64: ...Specifications...
Страница 65: ......
Страница 75: ...Specifications 3 10 1735 HD High Definition Waveform Monitor...
Страница 77: ......
Страница 78: ...Theory of Operation...
Страница 79: ......
Страница 96: ...Theory of Operation 1735 HD High Definition Waveform Monitor 4 17 Figure 4 2 1735 HD Clamp pulse generator...
Страница 109: ...Theory of Operation 4 30 1735 HD High Definition Waveform Monitor...
Страница 110: ...Performance Verification...
Страница 111: ......
Страница 134: ...Adjustment Procedure...
Страница 135: ......
Страница 158: ...Maintenance...
Страница 159: ......
Страница 163: ...Preventive Maintenance 7 4 1735 HD High Definition Waveform Monitor...
Страница 168: ...Troubleshooting 1735 HD High Definition Waveform Monitor 7 9 Figure 7 3 Multiple pin connectors...
Страница 179: ...Corrective Maintenance 7 20 1735 HD High Definition Waveform Monitor...
Страница 181: ...Repackaging 7 22 1735 HD High Definition Waveform Monitor Figure 7 10 Repackaging...
Страница 182: ...Replaceable Electrical Parts...
Страница 183: ......
Страница 252: ...Diagrams...
Страница 253: ......
Страница 256: ...Replaceable Mechanical Parts...
Страница 257: ......
Страница 265: ...Replaceable Mechanical Parts 10 8 1735 HD High Definition Waveform Monitor...
Страница 266: ...Appendices...
Страница 267: ......
Страница 270: ......
Страница 271: ......