Electrical Ratings (Maximum):
NC150N/NC200N Master Station ................. 24 VDC, .36 amps
PK151A Power & Control Unit ......... 24 VAC, 60 Hz, 1.25 amps
PK152 Power & Control Unit ........... 24 VAC, 60 Hz, 1.25 amps
IR150B Staff Station ..................................... 24 VDC, .09 amps
IR151B Patient Station ................................. 24 VDC, .09 amps
IR152B Patient Station ................................. 24 VDC, .09 amps
IR155B Patient Station ................................. 24 VDC, .09 amps
SF155B Emergency Station ......................... 24 VDC, .03 amps
SF156B Code Station ................................... 24 VDC, .06 amps
LI150B Duty Station...................................... 24 VDC, .03 amps
LI381 Corridor Light ...................................... 24 VDC, .06 amps
LI382 Corridor Zone Light........................... 24 VDC, .068 amps
LI382LED Corridor Zone Light...................... 24 VDC, .08 amps
PK800A Secondary Power Supply .................... 24 VAC, 60 Hz
Wiring Notes:
1. Selector switch shown is representative of both NC150N and NC200N
Master Stations.
2. Terminal identifications shown are not in actual order.
3. Main supply circuit: Connect to 117 VAC, 60 Hz (40 watts max). Use
18 gauge wire minimum. SS106 Transformer rating is 24 VAC, 30 VA max.
4. Use shielded cable if not in a metal conduit. If shielded cable is used, shield
drain wire replaces conductor connected to Terminal C on
talk
/
listen
switch
module.
5. All wiring is #18 gauge unless otherwise specified.
6. System wire description:
C = Audio common
R = Normal call
P = +24 VDC
Q = Emergency common
X = Audio selective
L = Lamp wire selective
Figure 9—NC150N, NC200N Nurse Call Systems
Notes and Electrical Ratings