System Installation
Installation Procedure
• Read the following instructions concerning system equipment and determine
installation methods before proceeding.
• Determine equipment locations.
• Install wiring.
• Install housing.
• Check wires.
• Connect equipment.
• Check connections.
Equipment Locations
NC150N, NC200N Master Stations: Locate NC150N or NC200N Master Station within
easy reach of operating personnel. Do not exceed operating temperature of 10–30°C.
IR150B Staff Stations: Locate IR150B stations where convenient to operate, but not in
the same room as the NC150N or NC200N Master Station. Do not exceed operating
temperature range of 10–40°C.
IR151B, IR152B, IR155B Patient Stations: Locate patient stations where convenient
for operation. Do not exceed operating temperature range of 10–40°C.
SF155B Emergency Switch: Locate emergency stations where convenient for operation.
Avoid areas where direct contact with water may occur. The SF155B includes a 6’
long pull cord permitting installation high enough to provide easy operation by the
nurse and by seated or prone patients. The SF155B may be used without the cord
as a pull-down actuated switch.
SF156B Code Switch: Locate code stations where convenient for operation. Avoid areas
where direct contact with water may occur.
LI150B Duty Stations: Locate duty stations as needed and where convenient for
operation. Location should provide for unobstructed visibility of the call indicator.
LI381 Corridor Lights: Locate corridor lights in the corridor above or beside the door
of the associated room. Location should provide for unobstructed visibility of the
corridor light in both directions.
LI382, LI382LED Corridor Zone Lights: Locate corridor zone lights in the corridor
area nearest the nurses central monitoring station. Location should provide for
unobstructed visibility of each corridor zone light from the central location.
SF301A, SF302 Call Cords: Insert call cord plugs into the associated station jacks as
indicated on the stations. Call cords provided are 6’ or 10’ in length.
PK151A, PK152 Power & Control Units: Locate the PK151A or PK152 and IH151N
Junction Box in an accessible area. Do not exceed operating temperature range of
10–30°C. Location should provide for convenient cable runs to remote and master
stations. Cable run from the PK151A or PK152 to the NC150N or NC200N Master
Station must not exceed 100’.
PK800A Secondary Power Supply: Locate the PK800A in an accessible area within 2
feet of the PK151A or PK152 that it is to be connected to. Do not exceed operating
temperature range of 10–30°C. The PK800A is for use in applications that exceed the
current load capacity of the PK151A or PK152 (1 amp). The PK800A allows the current
load to be increased to 3 amps. The FZ151 Fuse Assembly must be located between the
PK800A and the PK151A or PK152. See
Figure 16
for actual FZ151 connection points.