Tegam PS-CAL-900 Скачать руководство пользователя страница 8

 

 

 

10 TEGAM WAY • GENEVA, OHIO 44041 440-466-6100 • FAX 440-466-6110 • www.tegam.com 

3-2 

Section III – Work Station Configuration  

 

The Resource String is a standard VISA resource string.  The Resource String defines how PS-Cal will 

communicate with the instrument.  Typically, instrument communication is done over the GPIB bus, but 

PS-Cal is able to communicate with instruments via Ethernet, USB and RS-232.  The VISA resource string 

is a strictly formatted string.  GPIB configured instruments are formatted as follows: 

 

     GPIB<Card Number>::<GPIB Address>::INSTR. 

 

Note: Refer to the National Instruments VISA documentation for configuration formats for Ethernet, RS-

232 and USB instruments. 

 

The second tab on the Station Configuration form also contains a list of global station variables.  PS-Cal 

uses the variables as global variables to set values, custom configurations and execution options.  

 

 

Figure 3.2 Station Variables

 

 

 

After the Station Configuration has been edited, the data can be saved back to the configuration  files 

by pressing the Save and Close button.   

 
 

Editing Calibration Data for Standards 

The exact data required for each standard varies depending on the Measurement Methodology being used 

and configuration of the test station.  PS-Cal has a set of forms to help users manage the calibration data 

associated with the station’s standards.   

 

 

Editing Power Sensor Data 

To edit calibration data files for a power sensor standard select Configuration -> Edit Standard Data -

> Power Sensor in the menuPS-Cal opens a file dialog box.  

 

Содержание PS-CAL-900

Страница 1: ...10 TEGAM WAY GENEVA OHIO 44041 440 466 6100 FAX 440 466 6110 www tegam com PS CAL Power Sensor Calibration Software Instruction Manual PN PS CAL 900 Publication Date February 2018 REV B...

Страница 2: ...meters handheld temperature calibrators thermometers humidity and temperature control devices and more TEGAM also repairs and calibrates test and measurement equipment formerly manufactured by Electr...

Страница 3: ...rement Methodology Selection 3 1 Station Configuration 3 1 Editing Calibration Data for Standards 3 2 Editing Power Sensor Data 3 2 Editing Power Splitter Data 3 4 Editing Thermistor Mount Data 3 5 IV...

Страница 4: ...amage to the equipment or other property WARNING statements identify conditions or practices that could result in personal injury or loss of life Terms as Marked on Equipment CAUTION indicates a perso...

Страница 5: ...to the extent that such activity is expressly permitted by applicable law notwithstanding this limitation b Rental You may not rent or lease the SOFTWARE PRODUCT c Software Transfer This software is...

Страница 6: ...mework or higher If necessary update the workstation Net installation at microsoft com Installing Proper VISA Libraries PS CAL is designed to work will any VISA library Use the manufacturer VISA insta...

Страница 7: ...e right lists all the standards configured on this station This list contains the name of the driver followed by the instance of the driver The instance of the driver is used to distinguish instrument...

Страница 8: ...struments The second tab on the Station Configuration form also contains a list of global station variables PS Cal uses the variables as global variables to set values custom configurations and execut...

Страница 9: ...ta file is stored for the sensor file you wish to edit or view select the file and then press Open Figure 3 3 Editing Power Sensor Data PS Cal loads the data into a Power Sensor Data Edit form This fo...

Страница 10: ...g Data files for the first time you can load a blank data file from the C PS Cal_V4 _Blank Templates directory see Creating Templates for more information Navigate to the directory where the data file...

Страница 11: ...and New Data The Load Data tab allows the user to load existing PS Cal formatted data files from the hard drive for editing The Import Data tab allows the user to import data from various formats and...

Страница 12: ...nk Templates directory Figure 4 1 Creating a Calibration Template The first step in creating a calibration template is to open a blank template In the PS CAL menu select File Load and navigate to the...

Страница 13: ...ement Methodology select it and press Continue NOTE For Cal Factor tests using the TEGAM 1806 x Type IV Power Meters or the 1830A RF Thermistor Power Meter select from the RFBridge_Multirun methodolog...

Страница 14: ...logies and drivers have been configured Loading a Calibration Template Figure 4 4 Load a Calibration Template To load a calibration template in the PS CAL menu select Calibration Template Load and nav...

Страница 15: ...this column and save the changes back to the template This will force PS Cal to use your uncertainties on all reports 2 OnLabel For many power sensors that have a calibration label the calibration tem...

Страница 16: ...Calibration Two default tabs are loaded with every power sensor The first is the UUT tab containing information about the Unit Under Test UUT also known as Device Under Test DUT and the work order Da...

Страница 17: ...er you have completed a calibration or at any point during the calibration you can save the calibration data to disk We recommend you save your calibration data often to prevent loss of data due to a...

Страница 18: ...nsor calibration label is ready to be printed To print the calibration label 6 Select the Print Label tab of the power sensor being calibrated Under this is a report viewer tool used to generate the l...

Страница 19: ...tion are ready to be printed The last tab on the form is the Cal Report tab which combines all the data from each tab into one single report Figure 4 9 Print Calibration Report To print the Calibratio...

Страница 20: ...sensor in the batch This process saves time because many power sensor calibration configuration steps are time consuming Figure 4 11 Batch Mode To use PS Cal in Batch Mode 1 Click on UUT in the menu...

Отзывы: