Tegam PS-CAL-900 Скачать руководство пользователя страница 17

 

 

 

10 TEGAM WAY • GENEVA, OHIO 44041 440-466-6100 • FAX 440-466-6110 • www.tegam.com 

4-6 

Section IV - Operating Instructions 

 

Under each test procedure tab, there is a button bar with a button for each of the steps required to 

complete that calibration step.  Like the tabs, the buttons are organized in step order from left to right.  

The exact buttons that appear on the form varies with each test methodology.  

 

 

Figure 4.8 Calibration GUI 

These buttons are enabled and disabled as required by the test procedure. A button remains disabled until 

necessary prerequisite steps are performed. For example, the Cal button is turned off until the Configure 

Station button has been pressed and the configure station operation has been completed.  Then PS-Cal 

turns on the Cal button and allows you to go to the next step in the calibration process.  

Another button to make a special note of is the Halt button.  This button is used to halt the application. 

For example, if for some reason something was not hooked up correctly or you just need to stop the 

program, simply press the Halt button and PS-Cal will stop executing the calibration steps and reset all 

the standards to a safe state.  

 

Saving Calibration Data 

After you have completed a calibration (or at any point during the calibration), you can save the 

calibration data to disk.  We recommend you save your calibration data often, to prevent loss of data due 

to a power outage or power failure.  

To save your calibration data 

1.

 

Select UUT -> Save As and navigate to the “C:\PS-Cal_V4\CalResults\” directory. 

2.

 

Enter the file name of the calibration data you are saving (we recommend you use a standard 
naming conventions of either <JobNumber>.xml or <AssetNumber><CalDate>.xml).  

3.

 

Press Save.  

Uploading Calibration Data 

If the sensor has an EEPROM (Electronic, Erasable Programmable Read Only memory), you can upload the 

calibration data to the sensor.   

To upload the calibration data: 

1.

 

Click the Upload Data tab.   

2.

 

Click the Download Data button. This brings up the data for viewing. You will also see the word 
“Ready!” indicating that the data is now available for upload to the EEPROM on the sensor. 

3.

 

Click the Save/Backup Sensor Data to File button. This makes sure that the data is saved, in 
case there is a malfunction with the sensor prior to uploading. 

4.

 

Select whether to upload the data for 

a.

 

Header 

b.

 

Cal Factors 

c.

 

Linearity   

5.

 

Click the Upload Cal Data button. A dialog box prompts you to cycle the power on the power sensor. 
 
Note: to test that the data uploaded correctly, after cycling power on the sensor, with the sensor still 
connected, click the Download Data button. 

Содержание PS-CAL-900

Страница 1: ...10 TEGAM WAY GENEVA OHIO 44041 440 466 6100 FAX 440 466 6110 www tegam com PS CAL Power Sensor Calibration Software Instruction Manual PN PS CAL 900 Publication Date February 2018 REV B...

Страница 2: ...meters handheld temperature calibrators thermometers humidity and temperature control devices and more TEGAM also repairs and calibrates test and measurement equipment formerly manufactured by Electr...

Страница 3: ...rement Methodology Selection 3 1 Station Configuration 3 1 Editing Calibration Data for Standards 3 2 Editing Power Sensor Data 3 2 Editing Power Splitter Data 3 4 Editing Thermistor Mount Data 3 5 IV...

Страница 4: ...amage to the equipment or other property WARNING statements identify conditions or practices that could result in personal injury or loss of life Terms as Marked on Equipment CAUTION indicates a perso...

Страница 5: ...to the extent that such activity is expressly permitted by applicable law notwithstanding this limitation b Rental You may not rent or lease the SOFTWARE PRODUCT c Software Transfer This software is...

Страница 6: ...mework or higher If necessary update the workstation Net installation at microsoft com Installing Proper VISA Libraries PS CAL is designed to work will any VISA library Use the manufacturer VISA insta...

Страница 7: ...e right lists all the standards configured on this station This list contains the name of the driver followed by the instance of the driver The instance of the driver is used to distinguish instrument...

Страница 8: ...struments The second tab on the Station Configuration form also contains a list of global station variables PS Cal uses the variables as global variables to set values custom configurations and execut...

Страница 9: ...ta file is stored for the sensor file you wish to edit or view select the file and then press Open Figure 3 3 Editing Power Sensor Data PS Cal loads the data into a Power Sensor Data Edit form This fo...

Страница 10: ...g Data files for the first time you can load a blank data file from the C PS Cal_V4 _Blank Templates directory see Creating Templates for more information Navigate to the directory where the data file...

Страница 11: ...and New Data The Load Data tab allows the user to load existing PS Cal formatted data files from the hard drive for editing The Import Data tab allows the user to import data from various formats and...

Страница 12: ...nk Templates directory Figure 4 1 Creating a Calibration Template The first step in creating a calibration template is to open a blank template In the PS CAL menu select File Load and navigate to the...

Страница 13: ...ement Methodology select it and press Continue NOTE For Cal Factor tests using the TEGAM 1806 x Type IV Power Meters or the 1830A RF Thermistor Power Meter select from the RFBridge_Multirun methodolog...

Страница 14: ...logies and drivers have been configured Loading a Calibration Template Figure 4 4 Load a Calibration Template To load a calibration template in the PS CAL menu select Calibration Template Load and nav...

Страница 15: ...this column and save the changes back to the template This will force PS Cal to use your uncertainties on all reports 2 OnLabel For many power sensors that have a calibration label the calibration tem...

Страница 16: ...Calibration Two default tabs are loaded with every power sensor The first is the UUT tab containing information about the Unit Under Test UUT also known as Device Under Test DUT and the work order Da...

Страница 17: ...er you have completed a calibration or at any point during the calibration you can save the calibration data to disk We recommend you save your calibration data often to prevent loss of data due to a...

Страница 18: ...nsor calibration label is ready to be printed To print the calibration label 6 Select the Print Label tab of the power sensor being calibrated Under this is a report viewer tool used to generate the l...

Страница 19: ...tion are ready to be printed The last tab on the form is the Cal Report tab which combines all the data from each tab into one single report Figure 4 9 Print Calibration Report To print the Calibratio...

Страница 20: ...sensor in the batch This process saves time because many power sensor calibration configuration steps are time consuming Figure 4 11 Batch Mode To use PS Cal in Batch Mode 1 Click on UUT in the menu...

Отзывы: