3-1
Section III – Work Station Configuration
Measurement Methodology Selection
PS-Cal allows for multiple power sensor calibration methodologies. When using with TEGAM hardware the
RFBridge_MultiRun methodologies should be used.
Station Configuration
PS-Cal maintains a station configuration file on the local hard drive storing a list of the standards used and
the VISA resource string / GPIB address of the standard.
To open the station configuration file, select Configuration -> Edit Configuration
Figure 3.1 Station Configuration
The Available Drivers list on the left hand side contains a list of all the standards this version of PS-Cal
supports. The Configured Drivers list on the right lists all the standards configured on this station. This
list contains the name of the driver followed by the instance of the driver. The instance of the driver is
used to distinguish instruments in configuration using two or more standards of the same type. For
example, if the station is configured with two TEGAM 1830A’s, the Configured Drivers list would contain
two configuration lines, “TEGAM 1830A” and “TEGAM 1830A”.
Note: Only the standards listed in the Configured Drivers list are available for use in PS-Cal. All
drivers must be configured before they are available in the template configuration wizards.
The Driver Details section of the form shows the configuration details for this instrument. Each driver
must be configured with an Instance and Resource String.
The Instance must contain a number that uniquely identifies the driver / instrument. Typically, the first
instrument used is 1 and any other instruments of the same type are 2, 3, 4, etc.