3
Attention:
Situations of alert are marked by the ATTENTION symbol illustrated above.
Attention signs identify conditions or actions that could cause damage to the equipment or
to property.
ATTENTION: The AED may not operate correctly if it is activated or stored in conditions that
differ from those set forth herein
ATTENTION: The AED is resistant and reliable in various conditions of use. Nonetheless, an
overly aggressive AED treatment can cause damage to the unit or the accessories, causing
the warranty to lapse. Check the AED and its accessories for any damage, on a regular
basis and based on the instructions.
ATTENTION: Before delivering a shock, it is important to disconnect the patient from other
electrical medical equipment, such as blood flow meters, that may not be equipped with
protection from defibrillators. Also make sure that the electrodes are not touching metal
objects such as bed frames or tensioners.
ATTENTION: Only remove the protective film from the cartridge when you are ready to use
the electrodes.
ATTENTION: Do not use or activate the AED before reading the Operator Manual for the
AED.
ATTENTION: Do not use the AED or connect it to other equipment. If the AED is used with
or connected to other equipment, check operation carefully prior to use.
ATTENTION: Treating or transporting the patient during the ECG analysis could cause
interference with the diagnostic process, which could prolong analysis times. If the AED
reads SCARICA CONSIGLIATA (SHOCK RECOMMENDED) during treatment or transport, stop
the device and keep the patient as still as possible for at least 15 seconds before pressing
the Scarica (shock) button, so that the AED can re-confirm the analysis of the heartbeat
ATTENTION: It is necessary to perform periodic checks on this device to ensure, among
other things, that the AED is not damaged in any way.
ATTENTION: The electrodes are a single-use accessory and must be replaced after every
use or if the electrodes are broken/compromised in any way. If you suspect any damage,
the electrodes need to be replaced immediately.
ATTENTION: Do not use training electrodes with this AED
ATTENTION: Be particularly careful with patients with pacemakers. To determine whether
there is a pacemaker, it is necessary to run a physical exam and study the patient's medical
history. In patients with pacemakers, the AED could have less sensitivity and may not
detect all of the rhythms required for defibrillation.
ATTENTION: If the electrodes adhere well to the patient's chest, the AED can analyse the
precise ECG and avoid burns to the skin. If, however, the electrodes overlap on the
patient's chest, they will not deliver the defibrillation energy correctly.