41
ECG Performance Analysis
Rhythm class
AHA-DB
MIT-DB
CU-DB
VF-DB
Total number of
sample size
VF-shockable:
TP
747 19 220 566 1552
VF-shockable:
FN
82 0 18 53 153
VF-shockable:
sensitivity(%)
90.1 100.0
92.4 91.4 91.0
(>90)
VT-shockable:
TP
- - - 394
394
VT-shockable:
FN
- - - 63
63
VT-shockable:
sensitivity(%)
- - - 86.2
86.2(>75)
Non-shockable:
FP
0 0 0 1 1
NSR, PVC, APB,
SVP, Asystole-Non-
shockable: TN
3075 17150
1254 6774 28249
Specificity(%) 100.0
100.0 100.0 >99.9 >99.9
(>99)
Database for ECG analysis
From the official database of the AHA (American Heart Association)
From the official database of MIT (Massachusetts institute Technology)
(CU-DB: The Creighton University Sustained Ventricular Arrhythmia Database,
VF-DB: MIT-BIH Malignant Ventricular Arrhythmia Database)
ECG rhythm to determine whether it is advisable to deliver a shock
Ventricular fibrillation with an amplitude that is greater than or equal to 0.2mV
Ventricular Tachycardia with a heart frequency greater than or equal to 160 bpm