33
TROUBLESHOOTING
WARNING: If you are not sure that the measurements are accurate, check the
patient's vital signs using other methods; then, make sure that the AED is working
correctly.
WARNING: To reduce the risk of electric shock, do not attempt to remove the cover
for any reason. No parts can be replaced by the operator and any work
carried out on the AED can only be performed by a qualified technician.
General
If the AED detects an error, the status indicator will read “X”. Contact qualified service
personnel or your local supplier. Before contacting qualified staff or your local supplier, make
sure you have observed the ambient conditions set forth in the manual, such as
temperature, humidity, altitude, et cetera.
Corrective Measures
Check the expiration date of the electrodes. Change the electrodes if they have reached
their expiration date. Check the expiration date and the duration of the battery life in standby.
Change the battery if it has reached its expiration date or its battery life in standby.
Below is a list of possible errors and suggestions for corrective measures.
If the status indicator does not yet read “X” or if you hear a warning message when the AED
is on or if, for any reason, you suspect that the AED is not working correctly, contact qualified
staff or your local supplier or TECNO-GAZ directly, to receive assistance. (info@TECNO-
GAZ.co.kr)
1. There is no response when the cover of the AED is opened.
One of the CPU modules may not be working. Inform qualified staff or your local
supplier to check and replace the CPU module.
The battery may not be installed or it may be discharged. If the battery is not installed,
do so (See chapter on Battery Operation). If the battery is discharged, change it.
(Consult the Maintenance section)
2. The audio signals do not sound during operation.
Do not use the AED; contact qualified staff or your local supplier.
3. You can hear the audio signals, but the voice is not working properly.
Re-connect the cable or change the speaker.
4. The device plays the voice message “Connettere gli elettrodi. Inserire saldamente
il connettore (Connect the electrodes. Attach the connector firmly."
Firmly re-connect the electrode connector to the electrode attachment or replace the
electrodes.
5. The action icon does not flash.
Do not use the AED, contact qualified staff or your local supplier.
6. You cannot hear the voice message very well.
Do not use the AED, contact qualified staff or your local supplier.
7. The battery status indicator does not display 3 bars even though the battery has
been replaced.
If the battery status indicator does not display 3 bars even though the battery has
been replaced, do not use the AED and contact qualified staff or your local supplier.