7 ALARMS
7.1 WARNINGS
An alarm in the system is a recall caused by system problems that are sent to medical personnel by
means of a VISUAL AND AUDIO MESSAGE, when the delivery pressure in the equipment is insufficient
or a device failure occurs that may compromise the safety of the device.
THE DEVICE DOES NOT MONITOR THE PATIENT'S PHYSIOLOGICAL PARAMETERS.
DURING SEDATION MONITOR THE PATIENT USING A PULSE OXIMETER WITH
CUSTOMISED ALARM THRESHOLDS.
AT START-UP, TO CHECK THE OPERATION OF THE ACOUSTIC SIGNAL, THE DEVICE
EMITS 3 BEEPS, IN CASE OF ANOMALY DO NOT USE THE DEVICE AND CONTACT
THE TECHNICAL SUPPORT.
The device warns the operator by means of WARNINGS if there are any anomalies, THESE
ANOMALIES DO NOT AFFECT THE SAFETY OF THE DEVICE.
The WARNING indication appears on the screen, accompanied by an acoustic alarm that repeats with 3
beeps and a pause.
If the device is connected to the mains, the BARLED (see par 5.2) changes to Yellow.
To continue using the device it is necessary to press "Reset", the WARNING screen and the BUZZER
SIGNAL are removed, a BLINKING ICON remains on the DISPLAY and the BARLED is yellow.
If the cause of the fault is removed, the icon on the display disappears.
Below is a list of warnings and possible remedial actions:
Warning
Behaviour
Possible cause
Possible solution
A1*
The device does not deliver gas, the
O
2
pressure measured is 0
Closed or empty O
2
cylinder
Open the O
2
cylinder
Replace the O
2
cylinder
A2*
The device does not deliver N
2
O, N
2
O
pressure measured equal to 0
Closed or empty
N
2
O cylinder
Open the N
2
O cylinder
Replace the N
2
O cylinder
A3*
Device only delivers O
2
, low O2
pressure measured (< 2 bar)
Possible problem
with the O
2
reducer
Open the O
2
cylinder
Replace the O
2
cylinder
Check the O
2
pressure
reducer **
A5*
The device does not deliver gas, high
O2 pressure measured (> 6 bar)
A4*
Device only delivers O
2
, low N
2
O
pressure measured (< 2 bar)
Possible problem
with N
2
O reducer
Open the N
2
O cylinder
Replace the N
2
O cylinder
Check the N
2
O pressure
reducer **
A6*
Device only delivers O
2
, high N
2
O
pressure measured (> 6 bar)
A7***
Battery < 20%
The device signals that the battery is
running low.
-
Connect the device to the
mains
A8
Battery
≤
0%
The device interrupts the cycle
Low battery
-
Connect the device to the
mains or switch off the
device to avoid damaging
the battery
* This warning appears only during sedation
** If the pressure relief valve on the reducer vents, the reducer must be replaced.
*** Only if a sedation is being performed