ENGLISH
25
8.2
Scheduled maintenance
Before each use make sure that the BAG, SPIRAL TUBE and MASK KIT are
intact.
If MASTER FLUX is not used for over a month carry out an operational
check.
MASTER FLUX
EVERY 24 MONTHS OF USE OR IN CASE OF FAULTY OPERATION CONTACT THE DEALER
TO PERFORM A CHECK ON THE DEVICE
(ALL PROCEDURES MUST BE CARRIED OUT EXCLUSIVELY AT THE TECNO-GAZ PREMISES)
Shipping must be prepared in suitable packaging only:
Disassemble the flowmeter box
- Clean the device (except for the pressure reducers)
- Put the parts in plastic bags
Disassemble the pressure reducers from the cylinders
- Put them in clean plastic bags.
Use the original packaging.
PRESSURE REDUCERS
FOLLOW THE INSTRUCTIONS PROVIDED IN THE MANUALS OF THE DEVICES, BEING
CAREFUL OF SCHEDULED AND UNSCHEDULED MAINTENANCE.
CHECK FOR LEAKS:
- Close
the
“FLOW” knob (A – figure A9)
.
- Open the valves on the Oxygen and Nitrous oxide cylinders and read the values
provided by the pressure gauges.
- Close the cylinder valves
- Wait ~ 5 minutes and read the new values displayed on the pressure gauges.
- If the values are the same, the device does not have any leaks.
- If the values differ it is necessary to CHECK THE DEVICE.
CHECK THE ONE-WAY VALVE ON THE MASK (Only
Art. 1500/AS and 1501/AS)
-
Turn the perforated ring nut on the valve
(31 – figure A4)
until the holes close.
-
Let air into the mask intake at a very low pressure, you must be able to perceive the
outfeed flow from the mask.
-
Let air into the gas evacuation tube at a very low pressure, you must not be able to
perceive any flow in the mask.
Use original parts and accessories only
Содержание Master-Flux
Страница 7: ...ITALIANO 5 UNITS MM 428 516 1151 Figura A1 231 900 200 max 400 500 1305 Figura A2 ...
Страница 12: ...ITALIANO 10 Art 1605A Art 1606A Art 1607A 17M 31M 28M 29M 32M Figura A5 ...
Страница 14: ...ITALIANO 12 6 2 Installazione master flux a parete art 1501 AS e 1601AS 12 2 11 9 10 13 13A 26 F Figura A7 ...
Страница 35: ...ENGLISH 5 UNITS MM 428 516 1151 Figure A1 231 900 200 max 400 500 1305 Figure A2 ...
Страница 40: ...ENGLISH 10 Art 1605A Art 1606A Art 1607A 17M 31M 28M 29M 32M Figure A5 ...
Страница 63: ...FRANÇAIS 5 UNITS MM 428 516 1151 Figure A1 231 900 200 max 400 500 1305 Figure A2 ...
Страница 68: ...FRANÇAIS 10 Art 1605A Art 1606A Art 1607A 17M 31M 28M 29M 32M Figure A5 ...
Страница 70: ...FRANÇAIS 12 6 2 Installation de master flux mural art 1501 AS et 1601AS 12 2 11 9 10 13 13A 26 F Figure A7 ...
Страница 91: ...DEUTSCH 5 UNITS MM 428 516 1151 Abbildung A1 231 900 200 max 400 500 1305 Abbildung A2 ...
Страница 96: ...DEUTSCH 10 Art 1605A Art 1606A Art 1607A 17M 31M 28M 29M 32M Abbildung A5 ...
Страница 115: ......