ENGLISH
16
6.4
Evacuation of exhaled gases:
To safeguard the work environment and the staff, the exhaled gases must be conveyed
outside of the work environment.
Two systems can be used to discharge the gases externally:
- Direct connection to the outside of the office
- Connection to a surgical aspirator.
It is important to evaluate:
- Whether the aspirator is suitable for the evacuation of the gases (
N
2
O and O
2
)
- The aspirator exhaust must be outdoors and ventilated.
Make sure it is compliant with regulations in force in the country of use.
Direct connection to the outside of the office
UNIT-MOUNTED MASTER FLUX - Art. 1500/AS and 1600AS
Screw the nozzles
(8 - figure A3)
onto the unit.
Art. 1500/AS
Turn the perforated ring nut on the valve
(31 - figure A4)
until the holes close.
Art. 1600AS
Remove the supplied valve
(31M -
figure A5)
.
Connect the mask circuit gas exhaust tube to the nozzle
(8 - figure A3)
.
Connect the gas evacuation spiral tube
(36 - figure A4)
to the nozzle
(8 – figure A3)
on the unit.
Connect the end of the spiral tube to the outside and make sure there is a final filter.
WALL-MOUNTED MASTER FLUX - Art. 1501/AS and 1601AS
Connect the gas exhaust tube to the mask circuit.
Connect the gas evacuation spiral tube
(36-37 - figure A4)
to the nozzle
(34 - figure A4).
Connect the end of the spiral tube to the outside and make sure there is a final filter
(37 - figure A4)
.
Содержание Master-Flux
Страница 7: ...ITALIANO 5 UNITS MM 428 516 1151 Figura A1 231 900 200 max 400 500 1305 Figura A2 ...
Страница 12: ...ITALIANO 10 Art 1605A Art 1606A Art 1607A 17M 31M 28M 29M 32M Figura A5 ...
Страница 14: ...ITALIANO 12 6 2 Installazione master flux a parete art 1501 AS e 1601AS 12 2 11 9 10 13 13A 26 F Figura A7 ...
Страница 35: ...ENGLISH 5 UNITS MM 428 516 1151 Figure A1 231 900 200 max 400 500 1305 Figure A2 ...
Страница 40: ...ENGLISH 10 Art 1605A Art 1606A Art 1607A 17M 31M 28M 29M 32M Figure A5 ...
Страница 63: ...FRANÇAIS 5 UNITS MM 428 516 1151 Figure A1 231 900 200 max 400 500 1305 Figure A2 ...
Страница 68: ...FRANÇAIS 10 Art 1605A Art 1606A Art 1607A 17M 31M 28M 29M 32M Figure A5 ...
Страница 70: ...FRANÇAIS 12 6 2 Installation de master flux mural art 1501 AS et 1601AS 12 2 11 9 10 13 13A 26 F Figure A7 ...
Страница 91: ...DEUTSCH 5 UNITS MM 428 516 1151 Abbildung A1 231 900 200 max 400 500 1305 Abbildung A2 ...
Страница 96: ...DEUTSCH 10 Art 1605A Art 1606A Art 1607A 17M 31M 28M 29M 32M Abbildung A5 ...
Страница 115: ......