Quick Guide
Maintenance
15
Splash Protection Shields
1.
Remove the two screws of the inner Splash Protection
shield
n
.
2.
Carefully, remove the shield
o
.
3.
Clean and disinfect.
4.
Loosen the two screws of the outer Splash Protection
shield
p
.
5.
Carefully, remove the shield
q
.
6.
Clean and disinfect.
7.
Carefully, mount the outer Splash Protection shield
r
.
8.
Tighten the two screws
p
.
9.
Carefully, mount the inner Splash Protection shield
s
.
10. Tighten the two screws
n
.
11. Carefully, slide the inner Splash Protection shield up and
down to check for collision with the Sensors
t
.
2
3
4
5
6
8
1
8
7
CAUTION
In case of collision with the Sensors, slightly bend
the upper and/or lower tongue
u
of the inner
Splash Protection shield.
Содержание GENESIS FE500