QUICK GUIDE

NGLISH

Safety Instructions

page

3

Operation - Daily Routines

page

7

Preliminary Work

page

7

Starting

page

7

Sample Loading

page

8

Sample Unloading

page

10

Shut Down

page

11

Maintenance

page

11

Retrieve Information

page

17

Refilling and Disposal

page

21

Error Handling

page

29

Index

page

32

Doc. No. 391 401
Version 1.0
June 2002

Содержание GENESIS FE500

Страница 1: ...es page 7 Preliminary Work page 7 Starting page 7 Sample Loading page 8 Sample Unloading page 10 Shut Down page 11 Maintenance page 11 Retrieve Information page 17 Refilling and Disposal page 21 Error...

Страница 2: ...History Doc ID Version Description Issued 391 401 V1 0 First issue June 2002 The information contained in this document is subject to change without notice 2002 Tecan Schweiz AG Switzerland all right...

Страница 3: ...nment Definition Operator Qualified person who uses the equipment for its intended purpose Field Service Engineer Authorized person responsible for installation initial start up of the in strument and...

Страница 4: ...formation of all partially processed samples will be lost There fore an Uninterrupted Power Supply is highly recommen ded Upon attempting to restart an interrupted program the comput er screen provid...

Страница 5: ...te and federal environmental health safety laws and regulations Operating Manual Location The Operator s Manual p n 391 400 shall be kept in proximity of the The Operator s Manual is also located in t...

Страница 6: ...tment of Communication FCC These limits are de signed to provide reasonable protection against harmful inter ference when equipment is operated in a commercial environment This equipment generates use...

Страница 7: ...he Chapter Refilling and Disposal on page 21 Make sure four Centrifuge Buckets are placed on the Worktable correctly oriented Check also the Buckets in the Centrifuge Check the orientation of the Buck...

Страница 8: ...hts the button for system stop Sample Loading Inserting Sample Tubes into the Loading Racks 2 WARNING Potential Biohazard Consider all clinical specimens surfaces or components that contain or have co...

Страница 9: ...Empty Loading Racks 5 Regularly remove the empty Racks q When space is no longer available a warning message is displayed asking for Racks to be removed CAUTION Always remove the trays after loading o...

Страница 10: ...f the interface buttons in front of the re placed Carrier until the red LED stops flashing WARNING Potential Biohazard Consider all clinical specimens surfaces or components that contain or have con t...

Страница 11: ...anged from red to orange to green 2 In the LOH menu select LW o Wait until the Appli cation has closed 3 Switch off the power p 4 Shut down the PC Maintenance Syringe and Diluter Valve 1 Check and tig...

Страница 12: ...rating air in the Tube or a leaking Diluter Valve Replace the Diluter Valve if necessary 5 DiTi Cone 1 Clean the DiTi Cone n with weak detergent and disinfect with alcohol 2 Tighten using the Cone Wre...

Страница 13: ...Interface 1 Clean and disinfect Unload Area Carriers and Racks n 2 Wipe down the Unload Interface o CAUTION Do not use detergent sprays to prevent mois ture from getting into the electronic system or...

Страница 14: ...Waste Station 1 If the Aliquoter arm is above the Waste Station move it to the side 2 Carefully remove the compartment n 3 Clean and disinfect 4 Reinsert the compartment o CAUTION Do not use detergen...

Страница 15: ...the shield q 6 Clean and disinfect 7 Carefully mount the outer Splash Protection shield r 8 Tighten the two screws p 9 Carefully mount the inner Splash Protection shield s 10 Tighten the two screws n...

Страница 16: ...Quick Guide Maintenance 16 Arm Guide Rails Clean the Arm Guide Rails n with a lint free tissue Centrifuge Clean and grease the bearing surfaces of the hangers Consult Operator s Manual if necessary 1...

Страница 17: ...rcode r Position information Carrier Rack Sample s Status of the tube t ORVH button u Current Step v Next Step w History table with the executed actions and error status Displaying Routing Information...

Страница 18: ...Tubes with unsuccess ful Order Request from the Laboratory In formation System Number of Orders that have been suc cessfully processed Number of Orders with at least one tube that had an error during...

Страница 19: ...for addi tional information 1 In the tree select the requested object Barcode ID Order ID Primary Tube ID Tests ordered Secondary Tube ID selected Tube Tube barcode What workgroup the tube is assigne...

Страница 20: ...es data Find Opens the Find dialog box Select the search criteria and click LQG 1H W Enter Order No Enter Primary Tube Barcode Enter Secondary Tube Barcode Searches above actual position Searches belo...

Страница 21: ...he drawer q Status display for the Liquid Waste r Status display for the System Liq uid Container s DiTi Waste no function yet t Cap Waste no function yet WARNING Keep all disposables clean Use cautio...

Страница 22: ...move the DiTi Carrier o 3 Remove the empty DiTi Trays p Replacing the empty DiTi trays with full ones 4 Put the right side of the Rack on the two bolts and move it to the right as far as it will go q...

Страница 23: ...refilled DiTi Racks and click 5HILOO 6HO 5DFNV In the graphics the selected Rack s are filled 8 Click RQH Front Side 1 Positioning Pins CAUTION Make sure the Carrier is flush down on the worktable the...

Страница 24: ...tton No tubes are available Refill Yellow button Unknown number of tubes in the slot Refilling is possible Grey button The corresponding box is not inserted n Secondary Tube Slot 1 o Secondary Tube Sl...

Страница 25: ...he Secondary Tube Refill Box q 5 Turn the Secondary Tube Box upside down and slide it over the Secondary Tube Refill Box r NOTE The Secondary Tube refill box is shaped as to prevent wrong orientation...

Страница 26: ...turning both boxes over s 7 Slowly pull on the two tongues to remove the Secondary Tube Refill Box t Shake slightly while pulling 8 Check the Secondary Tubes Refill Box and the Secondary Tube Box for...

Страница 27: ...the label section Replace the label roll bracket and tighten the screw o 7 Insert the Barcode Label strip into the roller guides follow ing the arrows If the label has torn and needs to be re threaded...

Страница 28: ...he Refill STL Labels dialog box is displayed 2 Open the Printer drawer 3 Open the lock mechanisms p 4 Replace the Transfer Ribbon Waste Roll o and the Trans fer Ribbon Roll n Observe the correct Ribbo...

Страница 29: ...utton color changes from red to orange to green 2 Refer to the Section Global Stop and Restart of the Operator s Manual p n 391 400 WARNING Potential Biohazard Consider all clinical specimens surfaces...

Страница 30: ...d an er ror box is displayed According to the error the box functions are enabled or disabled For additional guidance click Click 5HWU If the error persists click Click the highlighted tabs blue and f...

Страница 31: ...Quick Guide Error Box 31 Notes...

Страница 32: ...check Diluter Valve 11 check Syringe 11 Chuck 14 DiTi Cone 12 Pick and Place Gripper 13 Printer 12 Splash Protection Shield 15 Tube Gripper 14 Unload Area 13 Unload Interface 13 Waste Station 14 P Pi...

Отзывы: