Quick Guide
Maintenance
12
Diluter Valve
4
Bubbles in the Tubes
q
after flushing may be caused by
untightened Syringe, untightened Plunger lock screw, lack
of System Liquid, aspirating air in the Tube, or a leaking
Diluter Valve. Replace the Diluter Valve, if necessary.
5
DiTi Cone
1
Clean the DiTi Cone
n
with weak detergent and disinfect
with alcohol.
2
Tighten using the Cone Wrench (Part No 619 517), if nec-
essary
o
.
DiTi Waste
Check DiTi Waste Bag and replace if necessary.
Printer
1
Check for lost Labels
n
.
2
Check and clean the Pressure Rollers
o
.
1
2
1
2