Quick Guide
Sample Unloading
10
Sample Unloading
L
1
Push one of the buttons in front of the Carrier to be un-
loaded
n
. Reset an erroneously pressed button by briefly
pressing it again.
2
When the
red
LED is flashing and the
green
LED is
dark, pull the Carrier carefully towards the front and re-
move it from the Worktable
o
.
3
Place a Carrier with empty Racks onto the
)(
Worktable. Observe the color strips on the Carrier front
and the Front Safety Panel
p
. Both LEDs are flashing
alternatively.
4
Push one of the interface buttons in front of the re-
placed Carrier until the
red
LED stops flashing.
WARNING: Potential Biohazard!
•
Consider all clinical specimens, surfaces
or components that contain or have con-
tacted human blood or serum as potential-
ly infectious. Wear gloves, lab coats, safety
glasses, and follow biosafety practices.
WARNING: Risk of Contamination!
•
Use caution handling the Carriers to avoid
sample splashing.
CAUTION
If the green LED is lit: Do not try to pull out the Car-
rier.
1
2
3
CAUTION
Do not remove the Carrier as long as the
green
LED
is lit.
CAUTION
To re-insert a not emptied Carrier:
Make sure the specimens are in their original spots
to avoid crashes and push the interface button
briefly.