RSA-G2 Getting Started Guide
Page 44
LN2 Controller Connection (optional)
1
Obtain the LN2 flex line from the LN2 accessory kit and connect between the LN2 Controller and a
liquid nitrogen supply source (bulk tank). Bulk tank needs to be a low pressure (22 psi) source.
2
Obtain the LN2 transfer line.
3
Connect the elbowed end of the transfer line to the LN2 Controller.
4
Connect the opposite end of the transfer line to the FCO LN2 supply connection.
5
Connect the power cable from the LN
2
Controller to the connector on the right side, above the
communications port.
Figure 25
FCO left- and right-hand configurations.