Axial Tests
The linear slide and stepper motor for independent vertical positioning of the instrument's
transducer enable the RSA-G2 to conduct additional axial tests for the ultimate testing
flexibility. A sample can be deformed under a constant linear strain rate, Hencky strain rate,
force, or stress for generating more traditional stress-strain curves.
The below left figure shows a Polyethylene film tested using the tension geometry at
25°C and a constant linear rate of 5 I-Im/s. A classical stress-strain curve is observed with
yield point. ultimate tensile strength, and fracture point.
----.
0
CL
�
'-"
en
en
�
+-
(/)
8
6
4
2
Constant Linear Rate on PE Storage Bag
Ultimate
Tensile
Strength
Fracture
Point
Temperature: 25°C
Rate: 5l-lm/s
Tension
o
10
20
30
40
Strain
(%)
50
60
70
80
Testing Modes and Applications
25
c
Q
.....
(/)
(f)
(f)
�
.....
(/)
Axial Test: Strain Rate Controlled
Linear rate
Hencky rate
Time
Axial Test: Force/Stress Controlled
Time
Содержание RSA-G2
Страница 1: ......
Страница 9: ...Forced Convection Oven with tension clamp Technology 8...
Страница 13: ...IMMERSION TESTING I CLAMPING SYSTEMS...
Страница 15: ...DETA I DIELECTRIC THERMAL ANALYSIS ACCESSORY DETA...
Страница 29: ...tainstruments com 2019 TA Instruments All rights reserved L50027 001...
Страница 30: ...Verso Filler Page z z...
Страница 31: ...n getting Started guide Revision D Issued July 2022...
Страница 99: ...Issued November 2013 Supplement to Getting Started Guide TA Instruments Manual Supplement...
Страница 100: ...Page 2 TA Instruments Manual Supplement This page is intentionally blank...
Страница 113: ...Page 1 Page 1 RSA G2 Solids Analyzer Revision G Issued May 2022 Site Preparation Guide...