RSA-G2
I
TECHNOLOGY
Transducer Mount
with Linear Slide
RSA-G2 Design Advantage
The RSA-G2 transducer is mounted to the instrument frame by a linear
slide and stepper motor allowing for independent vertical positioning.
Transducer motion is via a precision ground lead screw attached to a
micro-stepping motor by a rigid preloaded duplex bearing eliminating
backlash. A linear optical encoder is mounted directly between the
stationary frame and moving bracket for precision head positioning to
an accuracy of 0.1 !-1m.
Benefits of the independent transducer
positioning are:
•
Ease of clamp installation and sample loading
•
Compensation of sample expansion/contraction during experiments
•
Additional testing capabilities to large deformations such as strain rate
tests to pull samples until failure
Technology
6
Linear Optical Encoder
Lead Screw
Duplex Bearing
Micro-Stepping Motor
Содержание RSA-G2
Страница 1: ......
Страница 9: ...Forced Convection Oven with tension clamp Technology 8...
Страница 13: ...IMMERSION TESTING I CLAMPING SYSTEMS...
Страница 15: ...DETA I DIELECTRIC THERMAL ANALYSIS ACCESSORY DETA...
Страница 29: ...tainstruments com 2019 TA Instruments All rights reserved L50027 001...
Страница 30: ...Verso Filler Page z z...
Страница 31: ...n getting Started guide Revision D Issued July 2022...
Страница 99: ...Issued November 2013 Supplement to Getting Started Guide TA Instruments Manual Supplement...
Страница 100: ...Page 2 TA Instruments Manual Supplement This page is intentionally blank...
Страница 113: ...Page 1 Page 1 RSA G2 Solids Analyzer Revision G Issued May 2022 Site Preparation Guide...