Handling technique
Locking Implants to Rods
Variation A: Place Sleeve and
Nut Consecutively
2
Place nut on implant
Use the socket wrench for twelve point nut, with L-handle
to pick up the nut from the loading station and screw it on
to the implant thread (screw or hook).
3
Tighten the nut finger-tight
Tighten the nut using the socket wrench for twelve point nut
with L-handle. The socket wrench for the countertorque
is spring loaded and can be pressed downwards continuously
with the left hand using the T-handle. To tighten the nut
further, lift the L-handle of the socket wrench with the right
hand and engage it again.
Note:
If using a 6 mm rod, a few threads will remain visible
on the nut.