Synthes
15
2
Preliminary fixation of the plate to the bone
Required instruments
Drill Bit Ø 2,5 mm, length 110/85 mm, for Quick Coupling
310.250
Universal Drill Guide 3.5 (2.5/3.5)
323.360
LCP Drill Sleeve 2.7 (head LCP 2.4), with scale up to 60 mm
323.061
Use a K-wire through the drill sleeve in the distal hole to
fix the distal plate position. Make sure no collision with the
already implemented screws occurs.
Use the drill guide and the drill bit
2.5 mm to pre-drill both
cortices. Insert a 3.5 mm cortex screw through the long hole
of the plate.
3
Fix the distal part of the plate to the bone
Use a similar procedure as for the dorsolateral plate to insert
the LCP locking or cortex screws (see dorsolateral plate with
support).
Important:
Careful drilling is necessary as collision with the
screws of the dorsolateral plate may occur. In case of collision
stop drilling and use adequate screw for fixation. Use other
available holes for application of more screws.
Important:
It is recommended to use minimum one screw on
the medial side and one screw on the lateral side which cross
the distal block. Screw length should be 40–60mm depending
on the size of the humerus.