Option: Positioning
and compression device PCD
LCP Distal Humerus Plates
1
Fix aiming block to the plate
Required instruments
Screwdriver Shaft Stardrive 3.5, T15
314.116
Fix the aiming block and PCD to the dorsolateral plate with
support.
2
Position dorsolateral plate
Required instruments
Aiming Block, left, for Aiming Arm No. 313.354, for DHP
313.351
Aiming Block, right, for Aiming Arm No. 313.354, for DHP
313.352
Drill Sleeve 2.7, for Aiming Arm No. 313.354, for DHP
313.353
Aiming Arm for DHP
313.354
Insert for Drill Sleeve 2.7
313.355
Position the dorsolateral plate in approximate position. The
point of bone contact on the medial side marks the exit
point of the screw inserted through the hole of the plate and
thus shows the exact direction of the screw.
Use K-wires through the aiming block for temporary fixation.
Positioning and Compression Device PCD with lengths
markings for guided positioning of the dorsolateral plate with
support.
The PCD with length markings (313.351–357) assists in finding
an optimal plate position to allow for the insertion of longest
possible screws across the distal articular block.