Synthes
19
3
Fix the distal part of the plate to the bone
Required instruments
Drill Bit
2.0 mm, with double markings
323.062
Holding Sleeve for LCP Screw Stardrive
2.4/2.7 mm
313.301
Screwdriver Shaft Stardrive, T8, cylindrical, with groove
313.304
Torque Limiter, 0.8 Nm, with AO/ASIF Quick Coupling
511.776
Use the 3 distal screws to fix the plate to the capitellum.
Use a similar procedure as for the dorsolateral plate to insert
the LCP or cortex screws.
The recommended screw length is 16–24 mm.
Important:
Be careful when inserting the screws for the capi-
tellum to not damage the joint surface due to too long
screws. It is recommended to check the position of the screws
with the image intensifier during movement of the elbow.
4
Continue the procedure
Continue the procedure according to “Medial plate”.