Synthes
23
LCP Distal Humerus Plates
Option: Positioning and compression device PCD
6
Insert additional screws
Required instruments
LCP Drill Sleeve 2.7 (head LCP 2.4), with scale up to 60 mm
323.061
Drill Bit
2.0 mm, with double markings
323.062
Screwdriver Shaft Stardrive, T8, cylindrical, with groove
313.304
Torque Limiter, 0.8 Nm, with AO/ASIF Quick Coupling
511.776
Use the LCP drill sleeve and the drill bit
2.0 mm to drill
additional holes.
Insert additional LCP locking or cortex screws as needed.
Important:
When inserting the screws into the capitellum,
be careful not to damage the joint due to the length of the
screws.
7
Continue the procedure
Continue the procedure according to “Dorsolateral plate with
support”
The PCD may stay in place and serve as an indicator for the
position of the screw in order to avoid collision with the
medial distal screws.
Note:
the PCD can also be used to position a 3.5 mm cortex
screws through the articular block.