60
Synthes
Expert Tibial Nail
3
Remove nail
Instrument
03.010.056
Combined Hammer, 700g
Extract the nail by applying gentle blows with the Hammer.
Alternative instrument
03.010.001
Extraction Screw for 516.100 Air Pulse
For extraction of intramedullary femoral
and tibial nails
Before removing the final locking screw, screw the extraction
screw into the nail and tighten it to prevent rotation or
displacement of the nail posteriorly below the tibial plateau.
Remove the remaining locking screw. Attach Air Pulse to
the extraction screw and extract the Expert Tibial Nail.
Implant removal
0X6.000.380_AC.qxp:0X6.000.380_AC 20.01.2009 10:31 Uhr Seite 60