a
b
c
d
34
Synthes
Expert Tibial Nail
6
Insert locking screw
Instruments
03.010.107
Screwdriver Stardrive, T25,
length 330 mm
03.010.112
Holding Sleeve, with Locking Device
Insert the appropriate length locking screw using the screw-
driver Stardrive T25 and the holding sleeve, if needed.
Verify screw length under image intensification. If needed,
a second locking screw may be inserted using the same tech-
nique.
Note:
In the event of diastasis, the backstroke technique can
be used after insertion of the second distal locking screw.
Alternatively the compression screw can be used, please refer
to step 26 (page 34).
Use the holding sleeve as described below:
a
Insert the holding sleeve onto the shaft of the screwdriver
and place the tip of the screwdriver in the recess of the
locking screw.
b
Push the holding sleeve in the direction of the locking
screw, the sleeve now holds the locking screw.
c
Lock the holding sleeve by tightening it counter clockwise.
d
Release the holding sleeve after insertion of the locking
screw by loosening it clockwise and pushing backwards.
Distal locking
0X6.000.380_AC.qxp:0X6.000.380_AC 20.01.2009 10:31 Uhr Seite 34