32
Synthes
Expert Tibial Nail
5
Determine the length of the locking screw
Instrument
03.010.106
Direct Measuring Device for Drill Bits
length 145 mm
Stop drilling immediately after both cortices and disassemble
the drill bit from the Radiolucent Drive. Ensure the correct
position of the drill bit beyond the far cortex. Place the direct
measuring device onto the drill bit. Read the graduation of
the measuring device at the end of the drill bit.
This corresponds to the appropriate locking screw length.
Distal locking
0X6.000.380_AC.qxp:0X6.000.380_AC 20.01.2009 10:31 Uhr Seite 32