Page 11 of 17
Receiver for Wireless Hand Control, EG1A and Wireless Hand Control, EG1A include RF transmitters
:
•
Frequency band of transmission: 2405MHz to 2475MHz
•
Type and frequency characteristics of the modulation: DSSS (direct sequence spread spectrum)/ QPSK
(offset quadrature phase shift keying)
•
Effective radiated power: 4dBm
FCC Compliance Statement
:
This device complies with part 15 of the FCC Rules. Operation is subject to the following two conditions: (1) This
device may not cause harmful interference, and (2) this device must accept any interference received, including
interference that may cause undesired operation.
ETL Authorization Mark
:
Technical Specifications for Devices Compliant with IEC-60601-1-2 4
th
Edition
The device complies with the following standards:
AAMI ES60601-1 Issued: 2006/03/09 Medical Electrical Equipment - Part 1: General Requirements for Basic Safety
and Essential Performance, Amendment 1
–
2012
CSA C22.2#60601-1 Issued: 2014/03/01 Ed: 3 Medical Electrical Equipment - Part 1: General Requirements for
Basic Safety and Essential Performance
IEC 60601-1-6 Issued: 2013/10/29 Ed: 3 Medical Electrical Equipment - Part 1-6: General Requirements for Basic
Safety and Essential Performance - Collateral Standard: Usability; Amendment 1
IEC 60601-1: 2005 + CORR. 1:2006 + CORR. 2:2007 + AM1:2012 (or IEC 60601-1: 2012 reprint) EN 60601-1:
2005 + A1:2013, IEC 60601-1 / EN 60601-1, Medical electrical equipment Part 1: General requirements for basic
safety and essential performance (CB Scheme)
IEC 60601-1-6:2010 (Third Edition) + A1:2013 for use in conjunction with IEC62366:2007 (First Edition) + A1:2014
and IEC 60601-1:2005 (Third Edition) + Corr.1 (2006) + Corr.2 (2007) + A1: 2012 or equivalent consolidated
version IEC 60601-1:2012 (Ad.3.1), Medical electrical equipment Part 1-6 General requirements for safety -
Collateral Standard: Usability
IEC 60601-1-2:2014 (Edition 4) Collateral Standard: Electromagnetic Compatibility- Medical Electrical Equipment.
.
Environmental Conditions
Wireless Hand Control, EG1A
(WIRELESS-HC)
Receiver for Wireless Hand Control,
EG1A (RECEIVER-HC)
Temperature
Operating: 18°-30°C (65°-85°F)
Operating: 18°-30°C (65°-85°F)
Transportation and Storage: -29°
–
+60°C
(-20° - +140°F)
Transportation and Storage: -29°
–
+60°C
(-20° - +140°F)
Relative Humidity
Operating: 30% to 70%
Operating: 30% to 70%
Transportation and Storage: 10% to 90%
Transportation and Storage: 50% to 90%