Recommended separation distances  

between portable and mobile RF communications equipment and the eG1 System

 

The eG1 System is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The 
customer or the user of the eG1 System can help prevent electromagnetic interference by maintaining a minimum distance between 
portable and mobile RF communications equipment (transmitters) and the eG1 System as recommended below, according to the 
maximum output power of the communications equipment.  

Rated maximum output power of 

transmitter 

 

Separation distance according to frequency of transmitter 

 

150 kHz to 80 MHz 

 

80 MHz to 800 MHz 

 

800 MHz to 2,5 GHz 

 

P

d

2

.

1

=

 

P

d

2

.

1

=

 

P

d

3

.

2

=

 

0,01  

0.12 

0.12 

0.23 

0,1  

0.38 

0.38 

0.73 

1  

1.2 

1.2 

2.3 

10  

3.8 

3.8 

7.3 

100  

12 

12 

23 

For transmitters rated at a maximum output power not listed above, the recommended separation distance 

in meters (m) can be 

estimated using the equation applicable to the frequency of the transmitter, where 

is the maximum output power rating of the 

transmitter in watts (W) according to the transmitter manufacturer.  

NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.  

NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from 
structures, objects and people.  

 

 

Customer Service 

 
Contact your local Depuy Synthes representative for Troubleshooting assistance.  
To request a paper copy of the IFU call 800-255-2500. 
 
 

Disposal of Waste 

 

Dispose of items contaminated with body fluids or with other biohazardous waste. Please send devices that are no 
longer used to your local DePuy Synthes representative

Devices must be disposed of as a healthcare medical 

device in accordance with hospital procedures. The device contains electronics/battery and should be disposed off 
in accordance to local and/or national regulations. 
  

 

 

 
 
 
 

 

 

The Anspach Effort, Inc 

4500 Riverside Drive 
Palm Beach Gardens, FL 33410 USA 

Tel: 

(800) 327 6887 / +1 561 627 1080

 

 

 

 

www.e-IFU.com 
US / 800-255-2500 
 

Содержание ANSPACH EG1

Страница 1: ...d for cutting and shaping bone including the cranium and spine Contraindications The eG1 High Speed System does not have any known product specific contraindications Potential Adverse Events As with a...

Страница 2: ...ction causing delay in surgery The Hand Control is provided sterile Remove product from the package using established operating room procedures for sterile transfer Store sterile devices in their orig...

Страница 3: ...its packaging is compromised Visually inspect device components for damage before using do not use if damage or wear is evident unrecognizable markings missing or removed part numbers corrosion etc M...

Страница 4: ...rical Equipment Portable RF communications equipment including peripherals such as antenna cables and external antennas should be used no closer than 30 cm 12 inches to any part of the eG1 system Othe...

Страница 5: ...e field Caution Do not push receiver connector into console connector port when out of alignment Figure 1 Receiver connects to the foot control connector port on console SC3001 or SC3002 3 Remove the...

Страница 6: ...dpiece and release pressure on button to decrease speed of handpiece Warning Keep hand control safety cover in locked position when not in use Figure 5 Hand control cover shown in unlocked position 6...

Страница 7: ...st All Instrument System components returned should be properly cleaned as applicable prior to shipping to the manufacturer Warning Transmissible Spongiform Encephalopathies TSE DePuy Synthes Power To...

Страница 8: ...ator will be solid blue Out of Range The blue light indicator will blink with a fast flash when device is out of range or the signal is lost due to interference Hand Control Battery is at FULL Capacit...

Страница 9: ...ure device maintain a minimum distance between portable and mobile RF communications equipment transmitters and the eG1 System as recommended in the Recommended separation distances between portable a...

Страница 10: ...mmunication Apparatus All Frequency Bands Concurrently with FCC Part 15 247 IC ICES 003 2016Ed 6 Information Technology Equipment Including Digital Apparatus Limits and Methods of Measurement Concurre...

Страница 11: ...e IEC 60601 1 6 Issued 2013 10 29 Ed 3 Medical Electrical Equipment Part 1 6 General Requirements for Basic Safety and Essential Performance Collateral Standard Usability Amendment 1 IEC 60601 1 2005...

Страница 12: ...tic establishments and those directly connected to the public low voltage power supply network which supplies buildings used for domestic purposes Harmonic emissions IEC 61000 3 2 Class A Voltage fluc...

Страница 13: ...r 25 periods 50Hz 30 periods 60Hz Interruptions 0 UT 100 dip in UT for 250 periods 50Hz 320 periods 60Hz Mains power quality should be that of a typical commercial or hospital environment If the user...

Страница 14: ...magnetic site survey a should be less than the compliance level in each frequency range b Interference may occur in the vicinity of equipment marked with the following symbol Radiated RF IEC 61000 4 3...

Страница 15: ...M 1900 DECT LTE Band 1 3 4 25 UMTS Pulse modulation b 217 Hz 2 0 3 28 1 845 1 970 2 450 2 400 2 570 Bluetooth WLAN 802 11 b g n RFID 2450 LTE Band 7 Pulse Modulation b 217 Hz 2 0 3 28 5 240 5 100 5 80...

Страница 16: ...stimated using the equation applicable to the frequency of the transmitter where P is the maximum output power rating of the transmitter in watts W according to the transmitter manufacturer NOTE 1 At...

Страница 17: ...DePuy Synthes Representative for more information All DePuy Synthes Implant Instructions for Use as well as general Instructions for Use are available as PDF files at www e ifu com www depuysynthes c...

Отзывы: