Page 4 of 17
of the hand control could result in reduced
performance, material degradation, and/or
patient contamination.
•
Do not immerse or sterilize Receiver for
Wireless Hand Control, EG1A.
•
Do not use in oxygen rich environment.
•
Dispose of items contaminated with body
fluids with other biohazardous waste.
•
At end of life recycle or dispose of device in
accordance with local and national
regulations.
•
Verify that the Wireless Hand Control, EG1A
is connected to the intended Receiver for
Wireless Hand Control, EG1A. If the receiver
controlling the console is paired to an
unintended hand control, disconnect the
receiver from the EG1 console to prevent
unintended motion of the handpiece.
hospitals (CISPR 11 class A). This equipment
generates, uses, and can radiate radio frequency
energy and, if not installed and used in accordance
with the instruction manual, may cause harmful
interference to radio communications.
•
Operation of this equipment in a residential area
(for which CISPR 11 class B is normally required)
this equipment might not offer adequate protection
to radio frequency communication services.
•
In the event of interference, the user might need to
take mitigation measures such as relocating or
reorienting the equipment.
•
Portable and Mobile Radio Frequency (RF)
communications equipment can affect Medical
Electrical Equipment.
•
Portable RF communications equipment (including
peripherals such as antenna cables and external
antennas) should be used no closer than 30 cm (12
inches) to any part of the eG1 system. Otherwise,
degradation of the performance of this equipment
could result, including intermittent loss of function
or inability to pair the device.
Precautions
•
United States Federal law restricts this
device to sale by or on order of a physician
or other licensed healthcare provider.
•
Do not use accessories other than those
provided by DePuy Synthes Power Tools
and specified for use with Anspach Systems.
•
To ensure equipment operates as designed,
read and follow manufacturer's instructions.
•
Use care to protect when handling, cleaning, and
during system use.
•
Do not push receiver connector into console
connector port when out of alignment.
Latex Information
Not made with natural rubber latex.
Wireless Hand Control Set Up
Prior to first use the equipment (EG1A) must be processed as per the included EG1 High Speed System Cleaning
and Sterilization instructions.
1. Inspect the equipment prior to use
Warning:
Wireless Hand Control System is only compatible with eG1 Handpiece (EG1A) and Console (SC3001,
SC3002). Use with other than the indicated high speed system may lead to injury or damage to product.
EG1A
Visually inspect for damage to the silicone hose or to the
electrical connector.
Connect to console and operate.
The handpiece should operate smoothly.
There is no requirement to operate with attachment or
dissection tool.